Philips Medical Systems Nederland B.v. Azurion 7 B12 recall: Philips has identified instances where X-ray imaging may not be initiated
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
The hazard: Device safety concern
This is a device enforcement record. Check whether the remedy is a correction, a software update or a removal before changing anything.
What the notice says: Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-1675-2026
- Published
- April 8, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Philips Medical Systems Nederland B.v.
- Product type
- Medical devices
- Units affected
- 670 units (106 US, 564 OUS)
- Distribution
- Nationwide
- Made in
- Netherlands
- Lot / code info
- (1) System Code: 722067; UDI: 00884838085350; System Serial Number: 221, 42, 172, 173, 109, 250, 103, 192, 92, 114, 225, 69, 214, 41, 98, 81, 123, 278, 46, 277, 115, 150, 163, 131, 71, 104, 132, 147, 208, 244, 75, 36, 234, 141, 70, 26, 148, 133, 57, 128, 188, 151, 68, 233, 111, 255, 43, 33, 22, 146, 256, 270, 238, 187, 74, 154, 157, 251, 237, 222, 95, 53, 262, 29, 59, 209, 260, 191, 274, 52, 107, 218, 169, 231, 142, 143, 144, 275, 127, 210, 211, 212, 213, 206, 30, 23, 265, 190, 268, 203, 138, 34, 113, 254, 140, 266, 60, 97, 45, 112, 134, 100, 101, 168, 235, 272, 58, 47, 273, 176, 64, 156, 276, 232, 54, 55, 165, 25, 271, 35, 202, 149, 243, 32, 224, 207, 76, 108, 226, 48, 264, 102, 84, 105, 99, 21, 159, 49, 185, 51, 239, 40, 94, 63, 96, 129, 245, 186, 130, 252, 161, 120, 170, 195, 196, 197, 171, 200, 193, 80, 28, 229, 145, 79, 93, 137, 50, 44, 38, 175, 227, 167, 78, 86, 91, 189, 198, 37, 72, 199, 88, 27, 121, 179, 183, 117, 118, 85, 89, 90, 119, 87, 31, 110, 178, 73, 194, 39, 56, 139, 61; (2) System Code: 722225; UDI: 00884838099265; System Serial Number: 104, 153, 46, 28, 72, 208, 41, 122, 147, 376, 262, 331, 112, 110, 207, 375, 404, 405, 68, 177, 291, 299, 400, 29, 141, 216, 217, 71, 37, 175, 205, 228, 229, 157, 340, 453, 53, 259, 260, 300, 268, 55, 235, 298, 233, 354, 289, 148, 249, 434, 435, 409, 84, 111, 227, 455, 365, 284, 362, 468, 56, 232, 230, 269, 345, 296, 462, 374, 246, 324, 128, 241, 180, 367, 431, 287, 450, 460, 152, 292, 145, 124, 475, 306, 469, 470, 437, 248, 338, 140, 465, 413, 457, 314, 83, 414, 433, 461, 426, 420, 421, 422, 168, 158, 290, 267, 454, 418, 386, 309, 310, 438, 257, 223, 366, 173, 43, 8, 146, 276, 301, 109, 5, 189, 200, 387, 451, 452, 364, 65, 66, 398, 196, 372, 293, 315, 254, 312, 343, 201, 320, 392, 38, 423, 14, 350, 243, 393, 373, 99, 206, 456, 4, 442, 327, 447, 96, 47, 48, 52, 446, 302, 195, 186, 448, 242, 167, 187, 255, 256, 385, 30, 384, 321, 459, 169, 134, 135, 79, 103, 214, 395, 427, 116, 439, 107, 477, 133, 313, 346, 476, 471, 472, 251, 253, 441, 326, 411, 368, 333, 252, 351, 283, 87, 349, 1, 123, 220, 184, 258, 237, 234, 218, 149, 231, 342, 396, 7, 159, 142, 144, 274, 178, 348, 17, 31, 226, 330, 114, 266, 219, 20, 32, 155, 121, 192, 117, 164, 82, 85, 194, 131, 132, 355, 222, 40, 353, 212, 106, 445, 370, 44, 317, 13, 430, 213, 54, 244, 245, 69, 319, 64, 165, 166, 224, 334, 18, 19, 77, 76, 280, 236, 161, 311, 10, 316, 150, 119, 88, 125, 190, 151, 281, 75, 211, 39, 127, 369, 60, 272, 193, 118, 341, 210, 86, 170, 171, 61, 332, 361, 115, 307, 221, 198, 35, 59, 36, 363, 154, 15, 62, 57, 335, 279, 191, 160, 275, 94, 95, 23, 139, 197, 402, 11, 391, 120, 58, 34, 21, 329, 322, 67, 278, 162, 89, 90, 92, 337, 91, 137, 108, 73, 93, 371, 179, 360, 143, 359, 303, 163, 286, 432, 379, 138, 325, 100, 247, 98, 16, 27, 80, 282, 42, 105, 377, 424, 74, 277, 63, 323, 240, 70, 12, 9, 204, 238; (3) System Code: 722235; UDI: 00884838116788; System Serial Number: 77, 127, 76, 88, 12, 43, 102, 59, 71, 11, 55, 57, 89, 13, 74, 105, 118, 125, 135, 41, 94, 70, 54, 72, 47, 4, 92, 139, 79, 93, 27, 78, 14, 20, 109, 108, 56, 106, 30, 31, 142, 107, 10, 3, 99, 137, 131, 82, 83, 84, 100, 126, 66, 67, 96, 63, 34, 28, 17, 18, 65, 40, 45, 44, 42, 52, 49, 101, 53, 48, 81, 103, 25, 24, 6, 38, 39, 7, 5, 22, 8, 110, 58, 111, 98, 138, 64, 9, 35, 19, 23, 21, 33;
- UPC codes
- 00884838085350, 00884838099265, 00884838116788
Products covered by this recall
Azurion 7 B12; System Code: (1)722067, (2)722225, (3)722235;
Check your product
The notice names 23 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 00884838085350
- 00884838099265
- 00884838116788
Serial number
On the rating plate, often next to the date of manufacture.
- 221
- 42
- 172
- 173
- 109
- 250
- 103
- 192
- 92
- 104
- 153
- 46
- 28
- 72
- 208
- 41
- 122
- 147
- 77
- 127
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Philips Medical Systems Nederland B.v. — This is a device enforcement record. Check whether the remedy is a correction, a software update or a removal before changing anything.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- Ask the supplier or clinic that provided it what the remedy is in your case.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
Replacements, repairs and reimbursement
Device remedies run through the supply chain, so the route depends on who sold you the unit.
- Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
- If the device was billed to insurance or Medicare, the replacement is usually billed the same way; you should not face a second deductible for a recall.
- Keep every piece of correspondence. If a recalled device caused harm, that paper trail is what a claim rests on.
- Consumer-purchased devices such as monitors and test kits are normally replaced free, with return shipping paid.
Company contact: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Veenpluis 6, Best
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Philips Medical Systems Nederland B.v. medical devices recalled?
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 00884838085350, 00884838099265, 00884838116788. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 2 months ago
Philips Medical Systems Nederland B.v. Allura Xper FD. Model Numbers: 722026 recall: Firmware issue potentially leading to loss of geometry movements
Firmware issue potentially leading to loss of geometry movements.
-
Class II · 2 months ago
Philips Azurion, Model Numbers: 722229, 722230, 722231, 722232 recall: Two software related issues: 1) Potential image quality degradation
Two software related issues: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate…
-
Class II · 2 months ago
Philips Azurion Rystem, Model Numbers: 722063, 722064, 722067 recall: Two software related issues affecting Azurion R2.2.10 Systems: 1) Potential
Two software related issues affecting Azurion R2.2.10 Systems: 1) Potential image quality degradation that may lead to delay of therapy, procedural…
-
Class II · 2 months ago
Philips Allura, Cardio Vascular-Allura Centron recall: Potential Loss of imaging functionality, loss of motorized movement
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
-
Class II · 2 months ago
Philips Azurion under the following System Descriptions recall: Potential Loss of imaging functionality, loss of motorized movement
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
-
Class II · 2 months ago
Philips Allura under the following System Descriptions recall: Potential Loss of imaging functionality, loss of motorized movement
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.