Class II U.S. Food and Drug Administration

Philips Medical Systems Nederland B.v. Allura Xper FD20 recall: Philips has identified that in a limited number of impacted systems the drip

Recall published 9 months ago

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

The hazard: Device safety concern

A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.

What the notice says: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.

Recall details

Recall number
Z-1066-2026
Published
January 21, 2026
Risk level
Class II
Status
Ongoing
Issued by
U.S. Food and Drug Administration
Brand
Philips Medical Systems Nederland B.v.
Product type
Medical devices
Units affected
143 units
Distribution
Nationwide
Made in
Netherlands
Lot / code info
Model Number: 722006; System Product Name: Allura Xper FD20; UDI-DI: N/A; Serial Numbers: 1370, 1376, 225, 249, 1273, 1496, 707, 1466, 1141, 1201, 359, 105, 285, 1433, 822, 1747, 318, 797, 124, 1518, 384, 1045, 328, 302, 1536, 514, 910, 2123, 1360, 1381, 149, 379, 387, 1733, 197, 314, 888, 189, 310, 209, 863, 325, 378, 2110, 166, 118, 1978, 304, 1696, 2524, 248, 1928, 294, 804, 1227, 56, 245, 1560, 1795, 902, 1295, 1123, 1424, 1958, 40, 1222, 1947, 2289, 1687, 2304, 666, 2086, 958, 1801, 2546, 1317, 1478, 589, 2154, 1557, 1580, 519, 273, 157, 1504, 836, 1365, 393, 2428, 1902, 2433, 1521, 115, 2023, 1003, 1509, 1494, 2171, 2391, 867, 1513, 29, 951, 1814, 284, 482, 455, 1763, 1899, 1968, 837, 856, 1537, 93, 1561, 1746, 1232, 430, CV.797, 1974, 507, 1949, 1275, 1427, 1967, 374, 116, CV.507, 7567201, 2225, 267, 931, 344, 745, 233, 506, 2230, 1864, 457, 280, 1556, 640, 2;

Products covered by this recall

Allura Xper FD20; Model Number: 722006;

Check your product

The notice names 9 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.

Model or item number

On the rating label, the box or the instruction sheet.

  • 722006

Serial number

On the rating plate, often next to the date of manufacture.

  • 1370
  • 1376
  • 225
  • 249
  • 1273
  • 1496
  • 707
  • 14

Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.

What to do about Device

Philips Medical Systems Nederland B.v. — A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.

  1. Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
  2. If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
  3. Ask the supplier or clinic that provided it what the remedy is in your case.
  4. If you are a facility, quarantine affected stock and log the correction.

Claiming on a recalled device

Device recalls are tracked by serial and lot number, so your claim moves faster if you lead with those.

  • Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.
  • Ask explicitly whether your unit is in the affected range; many recalls cover only part of a model line.
  • If you need the device daily, ask what interim arrangement they offer while the remedy is carried out.
  • Log the date, the name and the reference number of every call. Device recalls can run for months.

Company contact: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Veenpluis 4-6, Best

The official notice

This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.

Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls

Questions people ask about this recall

Why was Philips Medical Systems Nederland B.v. medical devices recalled?

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown. Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or…

How do I know if I have the recalled product?

Match the codes printed on your package against the notice: model number 722006. If the code on your package is not on that list, your unit is outside the recall.

What should I do with it right now?

Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.

Can I get my money back?

Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.

Where was the recalled product sold?

It was distributed nationwide.

Related recalls