Philips Medical Systems Nederland B.v. Allura Xper FD20/10 recall: Philips has identified instances where X-ray imaging may not be initiated
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
The hazard: Device safety concern
The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
What the notice says: Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-1668-2026
- Published
- April 8, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Philips Medical Systems Nederland B.v.
- Product type
- Medical devices
- Units affected
- 85 units (20 US, 65 OUS)
- Distribution
- Nationwide
- Made in
- Netherlands
- Lot / code info
- System Code: 722029; UDI: 00884838059078; System Serial Number: 57, 7, 88, 56, 126, 94, 30, 31, 21, 97, 91, 110, 100, 63, 60, 55, 62, 129, 71, 36, 26, 49, 114, 128, 108, 103, 107, 81, 43, 52, 119, 118, 80, 78, 47, 106, 123, 96, 122, 127, 14, 101, 116, 102, 115, 19, 109, 120, 92, 69, 27, 46, 93, 50, 51, 124, 87, 20, 32, 104, 34, 45, 59, 79, 24, 70, 39, 66, 82, 75, 40, 33, 23, 25, 85, 37, 44, 58, 38, 61, 98, 53, 89, 35, 48;
- UPC codes
- 00884838059078
Products covered by this recall
Allura Xper FD20/10; System Code: 722029;
Check your product
The notice names 10 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 00884838059078
Serial number
On the rating plate, often next to the date of manufacture.
- 57
- 88
- 56
- 126
- 94
- 30
- 31
- 21
- 97
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Philips Medical Systems Nederland B.v. — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- Report device problems to the FDA through MedWatch.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
- Ask the supplier or clinic that provided it what the remedy is in your case.
Who pays for a device recall
The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.
- If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
- Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
- If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
- Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.
Company contact: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Veenpluis 6, Best
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Philips Medical Systems Nederland B.v. medical devices recalled?
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 00884838059078. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 2 months ago
Philips Medical Systems Nederland B.v. Allura Xper FD. Model Numbers: 722026 recall: Firmware issue potentially leading to loss of geometry movements
Firmware issue potentially leading to loss of geometry movements.
-
Class II · 2 months ago
Philips Azurion, Model Numbers: 722229, 722230, 722231, 722232 recall: Two software related issues: 1) Potential image quality degradation
Two software related issues: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate…
-
Class II · 2 months ago
Philips Azurion Rystem, Model Numbers: 722063, 722064, 722067 recall: Two software related issues affecting Azurion R2.2.10 Systems: 1) Potential
Two software related issues affecting Azurion R2.2.10 Systems: 1) Potential image quality degradation that may lead to delay of therapy, procedural…
-
Class II · 2 months ago
Philips Allura, Cardio Vascular-Allura Centron recall: Potential Loss of imaging functionality, loss of motorized movement
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
-
Class II · 2 months ago
Philips Azurion under the following System Descriptions recall: Potential Loss of imaging functionality, loss of motorized movement
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
-
Class II · 2 months ago
Philips Allura under the following System Descriptions recall: Potential Loss of imaging functionality, loss of motorized movement
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.