Class II U.S. Food and Drug Administration

Philips Azurion Systems R3.0. Includes the following label descriptions recall: Nine (9) identified software issues which may result in the following: loss

Recall published 9 months ago

Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after start-up, misalignment of Marker Tool Overlay, longitudinal position error.

The hazard: Device safety concern

The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.

What the notice says: Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after start-up, misalignment of Marker Tool Overlay, longitudinal position error.

Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.

Recall details

Recall number
Z-1079-2026
Published
January 21, 2026
Risk level
Class II
Status
Ongoing
Issued by
U.S. Food and Drug Administration
Brand
Philips Medical Systems Nederland B.v.
Product type
Medical devices
Units affected
990 units
Distribution
Nationwide
Made in
Netherlands
Lot / code info
Software Version: R3.0. 1. Model Number: 722229. UDI Numbers: (01)00884838116726(21)22, (01)00884838116726(21)15, (01)00884838116726(21)16, (01)00884838116726(21)6, (01)00884838116726(21)8, (01)00884838116726(21)4, (01)00884838116726(21)24, (01)00884838116726(21)26, (01)00884838116726(21)20, (01)00884838116726(21)25, (01)00884838116726(21)11, (01)00884838116726(21)21, (01)00884838116726(21)13, (01)00884838116726(21)7, (01)00884838116726(21)10, (01)00884838116726(21)2, (01)00884838116726(21)3, (01)00884838116726(21)29. 2. Model Number: 722230. UDI Numbers: (01)00884838116733(21)2, (01)00884838116733(21)40, (01)00884838116733(21)9, (01)00884838116733(21)1, (01)00884838116733(21)38, (01)00884838116733(21)10, (01)00884838116733(21)45, (01)00884838116733(21)7, (01)00884838116733(21)27, (01)00884838116733(21)28, (01)00884838116733(21)39, (01)00884838116733(21)4, (01)00884838116733(21)5, (01)00884838116733(21)23, (01)00884838116733(21)41, (01)00884838116733(21)20, (01)00884838116733(21)19, (01)00884838116733(21)13, (01)00884838116733(21)15, (01)00884838116733(21)12, (01)00884838116733(21)22. 3. Model Number: 722231. UDI Numbers: (01)00884838116740(21)13, (01)00884838116740(21)29, (01)00884838116740(21)50, (01)00884838116740(21)21, (01)00884838116740(21)19, (01)00884838116740(21)12, (01)00884838116740(21)25, (01)00884838116740(21)4, (01)00884838116740(21)1, (01)00884838116740(21)5, (01)00884838116740(21)2, (01)00884838116740(21)73, (01)00884838116740(21)72, (01)00884838116740(21)75, (01)00884838116740(21)33, (01)00884838116740(21)47, (01)00884838116740(21)60, (01)00884838116740(21)53, (01)00884838116740(21)45, (01)00884838116740(21)61, (01)00884838116740(21)34, (01)00884838116740(21)54, (01)00884838116740(21)35, (01)00884838116740(21)49, (01)00884838116740(21)32, (01)00884838116740(21)28, (01)00884838116740(21)27, (01)00884838116740(21)26, (01)00884838116740(21)51, (01)00884838116740(21)11, (01)00884838116740(21)20, (01)00884838116740(21)41, (01)00884838116740(21)52, (01)00884838116740(21)3. 4. Model Number: 722232. UDI Numbers: (01)00884838116757(21)61, (01)00884838116757(21)60, (01)00884838116757(21)5, (01)00884838116757(21)49, (01)00884838116757(21)2, (01)00884838116757(21)91, (01)00884838116757(21)33, (01)00884838116757(21)34, (01)00884838116757(21)48, (01)00884838116757(21)46, (01)00884838116757(21)23, (01)00884838116757(21)43, (01)00884838116757(21)36, (01)00884838116757(21)96, (01)00884838116757(21)95, (01)00884838116757(21)10, (01)00884838116757(21)76, (01)00884838116757(21)42, (01)00884838116757(21)6, (01)00884838116757(21)77, (01)00884838116757(21)3, (01)00884838116757(21)1, (01)00884838116757(21)56, (01)00884838116757(21)51, (01)00884838116757(21)44, (01)00884838116757(21)30, (01)00884838116757(21)29, (01)00884838116757(21)28, (01)00884838116757(21)102, (01)00884838116757(21)19, (01)00884838116757(21)8, (01)00884838116757(21)81, (01)00884838116757(21)20, (01)00884838116757(21)16, (01)00884838116757(21)11, (01)00884838116757(21)112, (01)00884838116757(21)63, (01)00884838116757(21)68, (01)00884838116757(21)32, (01)00884838116757(21)35, (01)00884838116757(21)122, (01)00884838116757(21)14, (01)00884838116757(21)82, (01)00884838116757(21)13, (01)00884838116757(21)50, (01)00884838116757(21)57, (01)00884838116757(21)58, (01)00884838116757(21)4, (01)00884838116757(21)105, (01)00884838116757(21)75, (01)00884838116757(21)121, (01)00884838116757(21)120, (01)00884838116757(21)71, (01)00884838116757(21)72, (01)00884838116757(21)73, (01)00884838116757(21)70, (01)00884838116757(21)69, (01)00884838116757(21)37. 5. Model Number: 722235. UDI Numbers: (01)00884838116788(21)88, (01)00884838116788(21)43, (01)00884838116788(21)12, (01)00884838116788(21)102, (01)00884838116788(21)127, (01)00884838116788(21)57, (01)00884838116788(21)55, (01)00884838116788(21)11, (01)00884838116788(21)76, (01)00884838116788(21)71, (01)00884838116788(21)59, (01)00884838116788(21)77, (01)00884838116788(21)89, (01)00884838116788(21)56, (01)00884838116788(21)99, (01)0…
UPC codes
00884838116726, 00884838116733, 00884838116740, 00884838116757, 00884838116788

Products covered by this recall

Philips Azurion Systems R3.0. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12. Model Number: 722229. 2. Azurion 3 M15. Model Number: 722230. 3. Azurion 5 M12. Model Number: 722231. 4. Azurion 5 M20. Model Number: 722232. 5. Azurion 7 B12. Model Number: 722235. 6. Azurion 7 B20. Model Number: 722236. 7. Azurion 7 M12. Model Number: 722233. 8. Azurion 7 M20. Model Number: 722234.

Check your product

The notice names 6 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.

UPC / barcode

Read the twelve digits under the barcode on the package.

  • 00884838116726
  • 00884838116733
  • 00884838116740
  • 00884838116757
  • 00884838116788

Model or item number

On the rating label, the box or the instruction sheet.

  • 722234. — Software Version

Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.

What to do about Device

Philips Medical Systems Nederland B.v. — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.

  1. Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
  2. Report device problems to the FDA through MedWatch.
  3. If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
  4. Ask the supplier or clinic that provided it what the remedy is in your case.

Who pays for a device recall

The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.

  • If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
  • Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
  • If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
  • Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.

Company contact: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Veenpluis 4-6, Best

The official notice

This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.

Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls

Questions people ask about this recall

Why was Philips Medical Systems Nederland B.v. medical devices recalled?

Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after start-up, misalignment of Marker Tool Overlay, longitudinal position error. Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system…

How do I know if I have the recalled product?

Match the codes printed on your package against the notice: UPC 00884838116726, 00884838116733, 00884838116740; model number 722234. — Software Version. If the code on your package is not on that list, your unit is outside the recall.

What should I do with it right now?

Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.

Can I get my money back?

If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.

Where was the recalled product sold?

It was distributed nationwide.