Philips Azurion systems not configured with an optional auxiliary pan recall: Potential safety issue with Philips Azurion systems
Potential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module.
The hazard: Device safety concern
This is a device enforcement record. Check whether the remedy is a correction, a software update or a removal before changing anything.
What the notice says: Potential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-2225-2026
- Published
- May 27, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Philips Medical Systems Nederland B.v.
- Product type
- Medical devices
- Units affected
- 5,537 units
- Distribution
- Nationwide
- Made in
- Netherlands
- Lot / code info
- 1. Model Number [UDI]: 722063 [(01)00884838085275(21)77, (01)00884838085275(21)47, (01)00884838085275(21)146, (01)00884838085275(21)102, (01)00884838085275(21)43, (01)00884838085275(21)2, (01)00884838085275(21)99, (01)00884838085275(21)28, (01)00884838085275(21)144, (01)00884838085275(21)62, (01)00884838085275(21)119, (01)00884838085275(21)61, (01)00884838085275(21)29, (01)00884838085275(21)40, (01)00884838085275(21)70, (01)00884838085275(21)120, (01)00884838085275(21)121, (01)00884838085275(21)110, (01)00884838085275(21)53, (01)00884838085275(21)103, (01)00884838085275(21)97, (01)00884838085275(21)101, (01)00884838085275(21)143, (01)00884838085275(21)116, (01)00884838085275(21)93, (01)00884838085275(21)141, (01)00884838085275(21)115, (01)00884838085275(21)137, (01)00884838085275(21)150, (01)00884838085275(21)128, (01)00884838085275(21)4, (01)00884838085275(21)50, (01)00884838085275(21)129, (01)00884838085275(21)78, (01)00884838085275(21)76, (01)00884838085275(21)69, (01)00884838085275(21)59, (01)00884838085275(21)92, (01)00884838085275(21)139, (01)00884838085275(21)130, (01)00884838085275(21)154, (01)00884838085275(21)75, (01)00884838085275(21)24, (01)00884838085275(21)94, (01)00884838085275(21)100, (01)00884838085275(21)26, (01)00884838085275(21)106, (01)00884838085275(21)37, (01)00884838085275(21)36, (01)00884838085275(21)27, (01)00884838085275(21)15, (01)00884838085275(21)135, (01)00884838085275(21)123, (01)00884838085275(21)87, (01)00884838085275(21)85, (01)00884838085275(21)89, (01)00884838085275(21)88, (01)00884838085275(21)86, (01)00884838085275(21)90, (01)00884838085275(21)91, (01)00884838085275(21)79, (01)00884838085275(21)133, (01)00884838085275(21)81, (01)00884838085275(21)52, (01)00884838085275(21)48, (01)00884838085275(21)148, (01)00884838085275(21)55]; 722229 [(01)00884838116726(21)7, (01)00884838116726(21)3, (01)00884838116726(21)10, (01)00884838116726(21)11, (01)00884838116726(21)21, (01)00884838116726(21)13]. 2. Model Number [UDI]: 722064 [(01)00884838085282(21)513, (01)00884838085282(21)61, (01)00884838085282(21)162, (01)00884838085282(21)161, (01)00884838085282(21)297, (01)00884838085282(21)169, (01)00884838085282(21)131, (01)00884838085282(21)479, (01)00884838085282(21)190, (01)00884838085282(21)359, (01)00884838085282(21)220, (01)00884838085282(21)375, (01)00884838085282(21)173, (01)00884838085282(21)256, (01)00884838085282(21)290, (01)00884838085282(21)502, (01)00884838085282(21)425, (01)00884838085282(21)89, (01)00884838085282(21)303, (01)00884838085282(21)130, (01)00884838085282(21)157, (01)00884838085282(21)60, (01)00884838085282(21)253, (01)00884838085282(21)280, (01)00884838085282(21)373, (01)00884838085282(21)417, (01)00884838085282(21)277, (01)00884838085282(21)102, (01)00884838085282(21)467, (01)00884838085282(21)461, (01)00884838085282(21)137, (01)00884838085282(21)84, (01)00884838085282(21)510, (01)00884838085282(21)49, (01)00884838085282(21)246, (01)00884838085282(21)282, (01)00884838085282(21)354, (01)00884838085282(21)374, (01)00884838085282(21)274, (01)00884838085282(21)127, (01)00884838085282(21)106, (01)00884838085282(21)329, (01)00884838085282(21)276, (01)00884838085282(21)58, (01)00884838085282(21)81, (01)00884838085282(21)200, (01)00884838085282(21)179, (01)00884838085282(21)249, (01)00884838085282(21)156, (01)00884838085282(21)380, (01)00884838085282(21)288, (01)00884838085282(21)231, (01)00884838085282(21)287, (01)00884838085282(21)150, (01)00884838085282(21)233, (01)00884838085282(21)230, (01)00884838085282(21)44, (01)00884838085282(21)176, (01)00884838085282(21)149, (01)00884838085282(21)197, (01)00884838085282(21)289, (01)00884838085282(21)184, (01)00884838085282(21)6, (01)00884838085282(21)448, (01)00884838085282(21)10, (01)00884838085282(21)82, (01)00884838085282(21)35, (01)00884838085282(21)278, (01)00884838085282(21)55, (01)00884838085282(21)39, (01)00884838085282(21)45, (01)00884838085282(21)66, (01)00884838085282(21)40, (01)00884838085282(21)125, (01)00884838085282(21)484, …
- UPC codes
- 00884838085275, 00884838116726, 00884838085282
Products covered by this recall
Philips Azurion systems not configured with an optional auxiliary pan handle. Includes the below product descriptions and corresponding model numbers. 1. Azurion 3 M12; Model Numbers: 722063, 722229. 2. Azurion 3 M15; Model Numbers: 722222, 722064, 722280, 722230. 3. Azurion 5 M12; Model Numbers: 722227, 722231. 4. Azurion 5 M20; Model Numbers: 722228, 722232, 722281. 5. Azurion 7 B12; Model Numbers: 722067, 722225, 722235. 6. Azurion 7 B20; Model Numbers: 722068, 722226, 722236. 7. Azurion 7 M12; Model Numbers: 722078, 722223, 722233. 8. Azurion 7 M20; Model Numbers: 722079, 722224, 722234, 722282.
Check your product
The notice names 19 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 00884838085275
- 00884838116726
- 00884838085282
Model or item number
On the rating label, the box or the instruction sheet.
- 722063
- 722222
- 722064
- 722280
- 722227
- 722228
- 722232
- 722067
- 722225
- 722068
- 722226
- 722078
- 722223
- 722079
- 722224
- 722234
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Philips Medical Systems Nederland B.v. — This is a device enforcement record. Check whether the remedy is a correction, a software update or a removal before changing anything.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- Ask the supplier or clinic that provided it what the remedy is in your case.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
Replacements, repairs and reimbursement
Device remedies run through the supply chain, so the route depends on who sold you the unit.
- Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
- If the device was billed to insurance or Medicare, the replacement is usually billed the same way; you should not face a second deductible for a recall.
- Keep every piece of correspondence. If a recalled device caused harm, that paper trail is what a claim rests on.
- Consumer-purchased devices such as monitors and test kits are normally replaced free, with return shipping paid.
Company contact: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Veenpluis 6, Best
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Philips Medical Systems Nederland B.v. medical devices recalled?
Potential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module. Potential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module.
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 00884838085275, 00884838116726, 00884838085282; model number 722063, 722222, 722064. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 2 months ago
Philips Medical Systems Nederland B.v. Allura Xper FD. Model Numbers: 722026 recall: Firmware issue potentially leading to loss of geometry movements
Firmware issue potentially leading to loss of geometry movements.
-
Class II · 2 months ago
Philips Azurion, Model Numbers: 722229, 722230, 722231, 722232 recall: Two software related issues: 1) Potential image quality degradation
Two software related issues: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate…
-
Class II · 2 months ago
Philips Azurion Rystem, Model Numbers: 722063, 722064, 722067 recall: Two software related issues affecting Azurion R2.2.10 Systems: 1) Potential
Two software related issues affecting Azurion R2.2.10 Systems: 1) Potential image quality degradation that may lead to delay of therapy, procedural…
-
Class II · 2 months ago
Philips Allura, Cardio Vascular-Allura Centron recall: Potential Loss of imaging functionality, loss of motorized movement
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
-
Class II · 2 months ago
Philips Azurion under the following System Descriptions recall: Potential Loss of imaging functionality, loss of motorized movement
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
-
Class II · 2 months ago
Philips Allura under the following System Descriptions recall: Potential Loss of imaging functionality, loss of motorized movement
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.