Class II U.S. Food and Drug Administration

Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer recall: Software issue: When programming the SynchroMed Pump with the A810 CP App

Recall published 5 months ago

Software issue: When programming the SynchroMed Pump with the A810 CP App in Flex Infusion Mode, the infusion steps should automatically be arranged in consecutive order, from earliest to latest. However, in rare cases, the steps may be displayed out of order. If this occurs, the programmed Flex Infusion schedule will not be delivered at the intended time, if at all.

The hazard: Device malfunction

This is a device recall, and in FDA language that usually means a correction rather than a removal: the device stays in use and the manufacturer fixes, patches or re-instructs.

What the notice says: Software issue: When programming the SynchroMed Pump with the A810 CP App in Flex Infusion Mode, the infusion steps should automatically be arranged in consecutive order, from earliest to latest. However, in rare cases, the steps may be displayed out of order. If this occurs, the programmed Flex Infusion schedule will not be delivered at the intended time, if at all.

Why it matters: The device may not perform as intended — a mechanical failure, a software defect, an inaccurate reading, or an alarm that does not sound. For a device someone depends on, the consequence of the failure is usually worse than the failure itself.

Symptoms and when to call a doctor

If the device supports a vital function, do not disable or remove it on your own judgement. Contact the clinician who prescribed or fitted it, and use the back-up arrangement they gave you. Go to an emergency department if the symptoms the device is meant to manage are getting worse, or if an implanted device gives an unexpected alarm or shock.

This page is general information, not medical advice. If you feel unwell, a clinician who can see you is the only person able to judge your case.

Recall details

Recall number
Z-2201-2026
Published
May 20, 2026
Risk level
Class II
Status
Ongoing
Issued by
U.S. Food and Drug Administration
Brand
Medtronic Neuromodulation
Product type
Medical devices
Units affected
10165 units
Made in
United States
Lot / code info
Software application (app) version 1.x and 2.x, loaded on the following units: 1. Model Number CT900A, GTIN 00643169890923, Serial Numbers: NPL0004445; 2. Model Number CT900B, GTIN 00643169890947, Serial Numbers: NPL1000042, NPL1000045, NPL1000046, NPL1000048, NPL1000050, NPL1000051, NPL1000053, NPL1000054, NPL1000059, NPL1000060, NPL1000062, NPL1000065, NPL1000069, NPL1000070, NPL1000071, NPL1000072, NPL1000073, NPL1000074, NPL1000075, NPL1000077, NPL1000081, NPL1000082, NPL1000083, NPL1000085, NPL1000086, NPL1000087, NPL1000090, NPL1000091, NPL1000093, NPL1000094, NPL1000098, NPL1000117, NPL1000133, NPL1000135, NPL1000137, NPL1000138, NPL1000151, NPL1000159, NPL1000160, NPL1000164, NPL1000166, NPL1000167, NPL1000169, NPL1000173, NPL1000174, NPL1000179, NPL1000180, NPL1000183, NPL1000185, NPL1000186, NPL1000187, NPL1000188, NPL1000191, NPL1000192, NPL1000193, NPL1000194, NPL1000195, NPL1000196, NPL1000230, NPL1000235, NPL1000237, NPL1000242, NPL1000244, NPL1000245, NPL1000246, NPL1000248, NPL1000249, NPL1000252, NPL1000254, NPL1000257, NPL1000258, NPL1000281, NPL1000292, NPL1000297, NPL1000301, NPL1000303, NPL1000307, NPL1000321, NPL1000322, NPL1000323, NPL1000325, NPL1000326, NPL1000332, NPL1000334, NPL1000343, NPL1000361, NPL1000362, NPL1000363, NPL1000364, NPL1000365, NPL1000366, NPL1000367, NPL1000369, NPL1000370, NPL1000381, NPL1000483, NPL1000486, NPL1000487, NPL1000501, NPL1000502, NPL1000505, NPL1000506, NPL1000507, NPL1000509, NPL1000510, NPL1000512, NPL1000515, NPL1000516, NPL1000518, NPL1000519, NPL1000528, NPL1000529, NPL1000539, NPL1000540, NPL1000555, NPL1000556, NPL1000558, NPL1000560, NPL1000561, NPL1000562, NPL1000563, NPL1000564, NPL1000571, NPL1000573, NPL1000574, NPL1000575, NPL1000576, NPL1000577, NPL1000578, NPL1000580, NPL1000585, NPL1000592, NPL1000603, NPL1000606, NPL1000612, NPL1000627, NPL1000630, NPL1000631, NPL1000632, NPL1000634, NPL1000637, NPL1000641, NPL1000652, NPL1000653, NPL1001194, NPL1001195, NPL1001201, NPL1001204, NPL1001205, NPL1001206, NPL1001207, NPL1001208, NPL1001210, NPL1001211, NPL1001212, NPL1001216, NPL1001223, NPL1001224, NPL1001225, NPL1001226, NPL1001228, NPL1001229, NPL1001230, NPL1001231, NPL1001235, NPL1001236, NPL1001237, NPL1001238, NPL1001240, NPL1001241, NPL1001243, NPL1001244, NPL1001250, NPL1001251, NPL1001254, NPL1001259, NPL1001264, NPL1001265, NPL1001272, NPL1001273, NPL1001274, NPL1001277, NPL1001282, NPL1001283, NPL1001290, NPL1001291, NPL1001306, NPL1001307, NPL1001313, NPL1001314, NPL1001328, NPL1001330, NPL1001332, NPL1001333, NPL1001334, NPL1001336, NPL1001337, NPL1001339, NPL1001341, NPL1001345, NPL1001349, NPL1001350, NPL1001351, NPL1001353, NPL1001354, NPL1001360, NPL1001361, NPL1001362, NPL1001366, NPL1001367, NPL1001369, NPL1001372, NPL1001375, NPL1001380, NPL1001381, NPL1001385, NPL1001386, NPL1001389, NPL1001398, NPL1001407, NPL1001408, NPL1001410, NPL1001412, NPL1001419, NPL1001421, NPL1001423, NPL1001430, NPL1001431, NPL1001434, NPL1001435, NPL1001436, NPL1001439, NPL1001445, NPL1001449, NPL1001458, NPL1001461, NPL1001464, NPL1001465, NPL1001468, NPL1001482, NPL1001483, NPL1001486, NPL1001499, NPL1001500, NPL1001505, NPL1001507, NPL1001510, NPL1001512, NPL1001513, NPL1001514, NPL1001518, NPL1001520, NPL1001524, NPL1001526, NPL1001531, NPL1001532, NPL1001537, NPL1001539, NPL1001540, NPL1001541, NPL1001542, NPL1001544, NPL1001545, NPL1001546, NPL1001547, NPL1001549, NPL1001550, NPL1001678, NPL1001689, NPL1001694, NPL1001695, NPL1001697, NPL1001699, NPL1001700, NPL1001701, NPL1001702, NPL1001705, NPL1001707, NPL1001708, NPL1001709, NPL1001710, NPL1001711, NPL1001716, NPL1001717, NPL1001723, NPL1001724, NPL1001727, NPL1001728, NPL1001739, NPL1001740, NPL1001741, NPL1001742, NPL1001743, NPL1001744, NPL1001745, NPL1001748, NPL1001749, NPL1001750, NPL1001754, NPL1001755, NPL1001756, NPL1001757, NPL1001759, NPL1001760, NPL1001761, NPL1001763, NPL1001764, NPL1001765, NPL1001766, NPL1001767, NPL1001769, NPL1001770, NPL1001773, NPL1001777, NPL1001784, NPL100…
UPC codes
00643169890923, 00643169890947

Products covered by this recall

Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application, loaded on the following devices: 1. PROG CT900A CLINICIAN TABLET US, Model Number CT900A; 2. PROG CT900B CLINICIAN TABLET EU -UK, Model Number CT900B; 3. PROG CT900C CLINICIAN TABLET UK, Model Number CT900C; 4. PROG CT900D CLINICIAN TABLET GLOBAL, Model Number CT900D; 5. PROG CT900E CLINICIAN TABLET GLBL SMSNG, Model Number CT900E; 6. PROG CT900F CLINICIAN TABLET GLBL SMSNG, Model Number CT900F; implanted programmable infusion pump

Check your product

The notice names 13 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.

UPC / barcode

Read the twelve digits under the barcode on the package.

  • 00643169890923
  • 00643169890947

Model or item number

On the rating label, the box or the instruction sheet.

  • CT900A
  • CT900B
  • CT900C
  • CT900D
  • CT900E
  • CT900F

Serial number

On the rating plate, often next to the date of manufacture.

  • NPL0004445
  • NPL1000042
  • NPL1000045
  • NPL1000046
  • NPL10

Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.

What to do about Malfunction

Medtronic Neuromodulation — This is a device recall, and in FDA language that usually means a correction rather than a removal: the device stays in use and the manufacturer fixes, patches or re-instructs.

  1. Match the model, catalogue, lot and serial numbers on your device against the identifiers listed on this page.
  2. Report device problems to the FDA through MedWatch — that database is where this recall came from.
  3. If the device is implanted, or you depend on it daily, speak to your clinician before changing anything about how you use it.
  4. Check whether the remedy is a software update, and if so confirm afterwards in the device menu that the version actually changed.

Replacements, repairs and reimbursement

Device remedies run through the supply chain, so the route depends on who sold you the unit.

  • Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
  • If the device was billed to insurance or Medicare, the replacement is usually billed the same way; you should not face a second deductible for a recall.
  • Keep every piece of correspondence. If a recalled device caused harm, that paper trail is what a claim rests on.
  • Consumer-purchased devices such as monitors and test kits are normally replaced free, with return shipping paid.

Company contact: Medtronic Neuromodulation — 7000 Central Ave Ne, Minneapolis, MN, 55432-3568

The official notice

This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.

Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls

Questions people ask about this recall

Why was Medtronic Neuromodulation medical devices recalled?

Software issue: When programming the SynchroMed Pump with the A810 CP App in Flex Infusion Mode, the infusion steps should automatically be arranged in consecutive order, from earliest to latest. However, in rare cases, the steps may be displayed out of order. If this occurs, the programmed Flex Infusion schedule will not be delivered at the intended time, if at all. Software issue: When programming the SynchroMed Pump with the A810 CP App in Flex Infusion Mode, the infusion steps should automatically be arranged in consecutive order, from earliest to latest. However, in rare cases, the steps…

How do I know if I have the recalled product?

Match the codes printed on your package against the notice: UPC 00643169890923, 00643169890947; model number CT900A, CT900B, CT900C. If the code on your package is not on that list, your unit is outside the recall.

What should I do with it right now?

Match the model, catalogue, lot and serial numbers on your device against the identifiers listed on this page.

Can I get my money back?

Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.