Medtronic MiniMed™ 600 series or 700 series insulin pumps recall: Device & Drug Safety - Defect
Device & Drug Safety - Defect
The hazard: Device safety concern
The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
What the notice says: Medtronic began investigating this issue upon receiving customer reports of shortened battery life and a comprehensive analysis found that pumps that have been dropped, bumped, or experienced physical impact even once may result in shortened battery life due to damage to internal electrical components.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Published
- October 4, 2024
- Risk level
- Elevated risk
- Issued by
- U.S. Food and Drug Administration
- Brand
- Medtronic
- Product type
- Medical devices
Products covered by this recall
MiniMed™ 600 series or 700 series insulin pumps
What the company announced
Customers are being notified of the importance of checking built-in alerts and alarms for battery status and to contact Medtronic for pump replacement if affected by this issue
GALWAY, Ireland - October 4, 2024 – Medtronic plc (NYSE: MDT), a global leader in healthcare technology, voluntarily issued a field action starting on July 31, 2024, notifying global customers of its MiniMed™ 600 series or 700 series insulin pumps to follow their pump’s built-in alerts and alarms for battery status and to contact Medtronic if they observe changes in the battery life of their pump. Medtronic began investigating this issue upon receiving customer reports of shortened battery life and a comprehensive analysis found that pumps that have been dropped, bumped, or experienced physical impact even once may result in shortened battery life due to damage to internal electrical components. This electrical issue also causes battery alerts to occur when less battery life remains than the User Guide states. This may result in the pump stopping insulin delivery significantly sooner than usually expected and could lead to health risks like hyperglycemia or diabetic ketoacidosis (DKA), potentially requiring medical intervention. Most recently, Medtronic has received 170 reports of hyperglycemia >400mg/dL and 11 reports of diabetic ketoacidosis from January 2023 to September 2024 in the US potentially related to this issue. The FDA has classified this action as a class I recall.
Medtronic issued an updated notification to customers stating the following:
Even a single drop of the pump could result in reduced battery life, either immediately after the drop, or over time, and the issue will continue even after replacing the battery.
The pump will still generate battery alerts and alarms, but these alerts and alarms may not be delivered with the usual amount of battery life remaining before the battery must be replaced to avoid the pump stopping insulin delivery. Be prepared to replace the battery as soon as the “Low Battery Pump” alarm sounds.
Users should always carry an extra set of new AA lithium or alkaline batteries or fully charged NiMH batteries should they need to replace the battery sooner than expected.
Users who notice a significant decrease in battery life from their usual experience, regardless of whether their pump has been dropped, bumped, or has experienced physical impact, should contact their local Medtronic team to determine whether a replacement pump is needed. Customers in the U.S. can call 1-800-378-2292 for 24-hour technical support. Customer outside the U.S. can click on this link to find the number for their local support team.
Users are advised to have a back-up therapy available in case it’s needed.
‘’Medtronic began proactively notifying impacted pump users in July because of the high priority we place on patient safety and understand the important role of pumps in the care and management of diabetes,” said Robert Vigersky, MD, Chief Medical Officer, Medtronic Diabetes. “Our team is here 24 hours a day, 7 days a week to help if you experience battery depletion or any other issue with your pump. Patient safety is our priority, as is continuity of therapy, which is why we made the decision to voluntarily notify patients and will replace pumps when needed.”
The original notification and the updated notification are available on the Medtronic Diabetes website.
About the Diabetes Business at Medtronic ( www.medtronicdiabetes.com ) Medtronic is working together with the global community to change the way people manage diabetes. The company aims to transform diabetes care by expanding access, integrating care and improving outcomes, so people living with diabetes can enjoy greater freedom and better health.
About Medtronic
Bold thinking. Bolder actions. We are Medtronic. Medtronic plc, headquartered in Galway, Ireland, is the leading global healthcare technology company that boldly attacks the most challenging health problems facing humanity by searching out and finding solutions. Our Mission — to alleviate pain, restore health, and extend life — unites a global team of 95,000+ passionate people across more than 150 countries. Our technologies and therapies treat 70 health conditions and include cardiac devices, surgical robotics, insulin pumps, surgical tools, patient monitoring systems, and more. Powered by our diverse knowledge, insatiable curiosity, and desire to help all those who need it, we deliver innovative technologies that transform the lives of two people every second, every hour, every day. Expect more from us as we empower insight-driven care, experiences that put people first, and better outcomes for our world. In everything we do, we are engineering the extraordinary. For more information on Medtronic, visit www.Medtronic.com and follow Medtronic on LinkedIn.
Any forward-looking statements are subject to risks and uncertainties such as those described in Medtronic's periodic reports on file with the Securities and Exchange Commission. Actual results may differ materially from anticipated results .
Reproduced from the company announcement posted on fda.gov. Links and contact forms in the original are not repeated here; the official notice below has them.
Check your product
The notice names 1 code you can read straight off the package. Type the one printed on yours and the list below narrows to it.
Package size
Match the net weight or the piece count on the front of the pack.
- 400 mg
Codes are lifted verbatim from the FDA — Company recall announcements notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Medtronic — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- Ask the supplier or clinic that provided it what the remedy is in your case.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
Who pays for a device recall
The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.
- If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
- Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
- If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
- Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.
Remedy named in the notice: Customers are being notified of the importance of checking built-in alerts and alarms for battery status and to contact Medtronic for pump replacement if affected by this issue
Company contact: Medtronic plc — Consumers: 1-800-378-2292 — Media: Janet Cho, Public Relations, +1 (818) 403-7028
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Company recall announcements record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Medtronic medical devices recalled?
Device & Drug Safety - Defect Medtronic began investigating this issue upon receiving customer reports of shortened battery life and a comprehensive analysis found that pumps that have been dropped, bumped, or experienced physical impact even once may result in shortened battery life due to damage to internal electrical components.
How do I know if I have the recalled product?
The notice does not publish package codes for this recall. Compare the product name, size and the purchase window against the official FDA — Company recall announcements record, and contact the company if you are unsure.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
The notice says: Customers are being notified of the importance of checking built-in alerts and alarms for battery status and to contact Medtronic for pump replacement if affected by this issue
Related recalls
-
Class II · 6 months ago
Medtronic Sphere-9 Catheter recall: Ventricular fibrillation (VF) occurred during radiofrequency (RF) therapy
Ventricular fibrillation (VF) occurred during radiofrequency (RF) therapy with the Medtronic Sphere-9 catheter in patients who have a Biotronik…
-
Class II · 9 months ago
Medtronic Inc Abre Venous Self-Expanding Stent System, Model numbers AB9G14120090 recall: Stents may be impacted by a Nitinol anomaly which causes a reduced stent
Stents may be impacted by a Nitinol anomaly which causes a reduced stent diameter when deployed and corresponding reduced radial force. Issue may…
-
Class II · 11 months ago
Medtronic, Inc. Aurora EV-ICD and Clinical EV-ICD, single chamber recall: There is a potential for delayed time to high-voltage (HV) therapy
There is a potential for delayed time to high-voltage (HV) therapy should a rare sequence of events occur.
-
Class II · 10 days ago
Howmedica Osteonics Corp. TS Universal Baseplate Finishing Tray, Triathlon recall: Reprocessing instructions provided in LSTPI-B specify a sterilization dry
Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is…
-
Class II · 10 days ago
Ferno-Washington Inc Power F2 Powered Fastening System, Power F2 recall: Firmware issue may result in intermittent loss or degradation of powered
Firmware issue may result in intermittent loss or degradation of powered functionality when paired with compatible stretcher. The issue may lead to…
-
Class II · 10 days ago
Howmedica Osteonics Corp. Triathlon Hinge Femoral Prep Tray recall: Reprocessing instructions provided in LSTPI-B specify a sterilization dry
Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is…