Class II U.S. Food and Drug Administration

Medline Medical Procedure Kits labeled as Clovis Hysteroscopy recall: During an internal review, Medline identified representative samples did

Recall published 4 months ago

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

The hazard: Device safety concern

This is a device enforcement record. Check whether the remedy is a correction, a software update or a removal before changing anything.

What the notice says: During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.

Recall details

Recall number
Z-2505-2026
Published
June 24, 2026
Risk level
Class II
Status
Ongoing
Issued by
U.S. Food and Drug Administration
Brand
Medline Industries, LP
Product type
Medical devices
Units affected
37324 units
Distribution
Nationwide
Made in
United States
Lot / code info
Medline Kit Number/SKU DYNJ61063A: UDI/DI 10193489640458 (each), 40193489640459 (case), Lot Number 21KBV571; Medline Kit Number/SKU DYNJ61063B: UDI/DI 10195327193478 (each), 40195327193479 (case), Lot Number 23EBU351; Medline Kit Number/SKU DYNJ61063B: UDI/DI 10195327193478 (each), 40195327193479 (case), Lot Number 23DBQ190; Medline Kit Number/SKU DYNJ61063B: UDI/DI 10195327193478 (each), 40195327193479 (case), Lot Number 23BBE364; Medline Kit Number/SKU DYNJ61063B: UDI/DI 10195327193478 (each), 40195327193479 (case), Lot Number 22IBW094; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 26CMK666; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 26BMG709; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 26AMI644; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 25KMD133; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 25JMA568; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 25IMF717; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 25HMJ595; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 25FMH149; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 25EME964; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 25DMG466; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 25CMH227; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 25CMC736; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 25BMB744; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 24LMB964; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 24JMB516; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 24JMA159; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 24EMI862; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 24DMG741; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 24CBP782; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 24CMI475; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 24BBU531; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 24BBB690; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 23LBH349; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 23JBP275; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 23IBU481; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 23GBW750; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 23GBB691; Medline Kit Number/SKU DYNJ50874C: UDI/DI 10195327024963 (each), 40195327024964 (case), Lot Number 22ABS574; Medline Kit Number/SKU DYNJ50874C: UDI/DI 10195327024963 (each), 40195327024964 (case), Lot Number 22ABI415; Medline Kit Number/SKU DYNJ50874C: UDI/DI 10195327024963 (each), 40195327024964 (case), Lot Number 21KBN568; Medline Kit Number/SKU DYNJ50874D: UDI/DI 10195327165383 (each), 40195327165384 (case), Lot Number 24CBV299; Medline Kit Number/SKU DYNJ50874D: UDI/DI 10195327165383 (each), 4019532716…
UPC codes
10193489640458, 40193489640459, 10195327193478, 40195327193479, 10195327451684, 40195327451685, 10195327024963, 40195327024964, 10195327165383, 40195327165384

Products covered by this recall

MEDLINE Medical Procedure Kits labeled as CLOVIS HYSTEROSCOPY, CSC-N PELVISCOPY, D AND C PACK, D&C HYSTER PACK, FLWR GYN PACK, FS LAP TUBAL LIGATION, HILLSDALE D&C / HYSTEROSCOPY, LAVH NEW BRAUNFELS,TVH PACK, etc (see recall documents for a full list of affected products)

Check your product

The notice names 35 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.

UPC / barcode

Read the twelve digits under the barcode on the package.

  • 10193489640458
  • 40193489640459
  • 10195327193478
  • 40195327193479
  • 10195327451684
  • 40195327451685
  • 10195327024963
  • 40195327024964
  • 10195327165383
  • 40195327165384

Lot or batch code

Usually ink-stamped near the seam, the cap or the bottom of the pack.

  • 21KBV571
  • 23EBU351
  • 23DBQ190
  • 23BBE364
  • 22IBW094
  • 26CMK666
  • 26BMG709
  • 26AMI644
  • 25KMD133
  • 25JMA568
  • 25IMF717
  • 25HMJ595
  • 25FMH149
  • 25EME964
  • 25DMG466
  • 25CMH227
  • 25CMC736
  • 25BMB744
  • 24LMB964
  • 24JMB516

Model or item number

On the rating label, the box or the instruction sheet.

  • DYNJ61063A
  • DYNJ61063B
  • DYNJ61063C
  • DYNJ50874C
  • DYNJ50874D

Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.

What to do about Device

Medline Industries, LP — This is a device enforcement record. Check whether the remedy is a correction, a software update or a removal before changing anything.

  1. Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
  2. Report device problems to the FDA through MedWatch.
  3. If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
  4. Ask the supplier or clinic that provided it what the remedy is in your case.

Replacements, repairs and reimbursement

Device remedies run through the supply chain, so the route depends on who sold you the unit.

  • Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
  • If the device was billed to insurance or Medicare, the replacement is usually billed the same way; you should not face a second deductible for a recall.
  • Keep every piece of correspondence. If a recalled device caused harm, that paper trail is what a claim rests on.
  • Consumer-purchased devices such as monitors and test kits are normally replaced free, with return shipping paid.

Company contact: Medline Industries, LP — 3 Lakes Dr, Northfield, IL, 60093-2753

The official notice

This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.

Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls

Questions people ask about this recall

Why was Medline Industries, LP medical devices recalled?

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification. During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached…

How do I know if I have the recalled product?

Match the codes printed on your package against the notice: UPC 10193489640458, 40193489640459, 10195327193478; lot code 21KBV571, 23EBU351, 23DBQ190; model number DYNJ61063A, DYNJ61063B, DYNJ61063C. If the code on your package is not on that list, your unit is outside the recall.

What should I do with it right now?

Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.

Can I get my money back?

Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.

Where was the recalled product sold?

It was distributed nationwide.