Medline Convenience Kits: 1) Spinal CDS recall: Medline has identified issues related to calibration of the equipment used
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
The hazard: Sterility or particulate concern
The concern here is contamination of a product that bypasses the body’s normal defences, which is why the threshold for recalling it is low.
What the notice says: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Why it matters: The product was made or packaged in conditions that cannot assure sterility, or visible particles were found in it. For anything injected, infused or used in the eye, that is a direct route for contamination into the body.
Symptoms and when to call a doctor
Watch for fever, chills, pain, redness, swelling or discharge at an injection site, and for a red, painful or suddenly blurred eye after using an eye product. Any of these needs same-day medical attention. Do not wait to see whether it settles, and tell the clinician which product and lot you used so it can be reported against the recall.
This page is general information, not medical advice. If you feel unwell, a clinician who can see you is the only person able to judge your case.
Recall details
- Recall number
- Z-1421-2026
- Published
- February 25, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Medline Industries, LP
- Product type
- Medical devices
- Units affected
- 117907 units
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- 1) CDS940087AG, UDI-DI: 10193489847260(each), 40193489847261(case), Lot Number: 21EBC900; 2) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 22GBE442; 3) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 22EBG548; 4) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 22EBF695; 5) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 22EBA274; 6) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 22BBY060; 7) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 21KBH163; 8) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 21JBF936; 9) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 21IBJ260; 10) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 21GBI183; 11) CDS980832Q, UDI-DI: 10193489788655(each), 40193489788656(case), Lot Number: 21FBS254; 12) CDS980832Q, UDI-DI: 10193489788655(each), 40193489788656(case), Lot Number: 21BBL223; 13) CDS980832R, UDI-DI: 10193489900972(each), 40193489900973(case), Lot Number: 21LBO219; 14) CDS980832R, UDI-DI: 10193489900972(each), 40193489900973(case), Lot Number: 21KBQ650; 15) CDS980832R, UDI-DI: 10193489900972(each), 40193489900973(case), Lot Number: 21KBB625; 16) CDS980832R, UDI-DI: 10193489900972(each), 40193489900973(case), Lot Number: 21IBV695; 17) CDS980832R, UDI-DI: 10193489900972(each), 40193489900973(case), Lot Number: 21IBU197; 18) CDS980832R, UDI-DI: 10193489900972(each), 40193489900973(case), Lot Number: 21GBR895; 19) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 23JBI054; 20) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 23GBP400; 21) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 23EBW653; 22) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 23DBR887; 23) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 23CBG614; 24) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 23BBK758; 25) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 23BBJ844; 26) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 22OBK666; 27) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 22OBI093; 28) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 22JBH355; 29) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 22HBS531; 30) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 22GBM598; 31) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 22GBL549; 32) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 22GBK859; 33) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 22GBK858; 34) CDS980833Q, UDI-DI: 10193489900989(each), 40193489900980(case), Lot Number: 21LBL968; 35) CDS980833Q, UDI-DI: 10193489900989(each), 40193489900980(case), Lot Number: 21LBC616; 36) CDS980833Q, UDI-DI: 10193489900989(each), 40193489900980(case), Lot Number: 21KBF393; 37) CDS980833Q, UDI-DI: 10193489900989(each), 40193489900980(case), Lot Number: 21JBO077; 38) CDS980833Q, UDI-DI: 10193489900989(each), 40193489900980(case), Lot Number: 21GBV804; 39) CDS980833Q, UDI-DI: 10193489900989(each), 40193489900980(case), Lot Number: 21GBB808; 40) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 23EBK295; 41) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 23CBT556; 42) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 23BBP326; 43) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 23ABH483; 44) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 23ABG351; 45) CDS980833R, UDI-DI: 10195327064501…
- UPC codes
- 10193489847260, 40193489847261, 10193489933758, 40193489933759, 10193489788655, 40193489788656, 10193489900972, 40193489900973, 10195327064495, 40195327064496, 10193489900989, 40193489900980, 10195327064501, 40195327064502
Products covered by this recall
Medline Convenience Kits: 1) SPINAL CDS, Model Number: CDS940087AG; 2) SPINAL CDS, Model Number: CDS940087AI; 3) TOTAL HIP CDS-LF, Model Number: CDS980832Q; 4) TOTAL HIP CDS-LF, Model Number: CDS980832R; 5) TOTAL HIP CDS-LF, Model Number: CDS980832S; 6) TOTAL KNEE CDS-LF, Model Number: CDS980833Q; 7) TOTAL KNEE CDS-LF, Model Number: CDS980833R; 8) TOTAL HIP PROCEDURE, Model Number: CDS980865V; 9) TOTAL HIP PROCEDURE, Model Number: CDS980865W; 10) TOTAL HIP PROCEDURE, Model Number: CDS980865X; 11) TOTAL HIP PROCEDURE, Model Number: CDS980865Y; 12) TOTAL KNEE CDS, Model Number: CDS981454K; 13) TOTAL KNEE CDS, Model Number: CDS981454L; 14) KNEE ARTHROSCOPY CDS, Model Number: CDS981502K; 15) TOTAL KNEE, Model Number: CDS981505O; 16) TOTAL KNEE, Model Number: CDS981505P; 17) TOTAL KNEE, Model Number: CDS981505Q; 18) TOTAL KNEE, Model Number: CDS981505R; 19) OR LAMINECTOMY CDS, Model Number: CDS981858F; 20) LAMI FUSION, Model Number: CDS982035W; 21) ARMC TOTAL KNEE CDS, Model Number: CDS982281O; 22) ARMC TOTAL KNEE CDS, Model Number: CDS982281P; 23) ARMC TOTAL KNEE CDS, Model Number: CDS982281S; 24) TOTAL HIP CDS, Model Number: CDS983036K; 25) TOTAL KNEE CDS, Model Number: CDS983042K; 26) LOWER EXTREMITY CDS, Model Number: CDS983049K; 27) ZALE SPINE CDS, Model Number: CDS983182G; 28) TOTAL KNEE CDS, Model Number: CDS983297M; 29) SITTING CERVICAL CDS, Model Number: CDS983480G; 30) SHOULDER CDS, Model Number: CDS983581K; 31) SHOULDER ARTHROSCOPY CDS, Model Number: CDS983838G; 32) TOTAL HIP CDS, Model Number: CDS983840I; 33) TOTAL JOINT CDS-LF, Model Number: CDS983863D; 34) SHOULDER ARTHROSCOPY CDS, Model Number: CDS983879C; 35) TOTAL JOINT, Model Number: CDS983903D; 36) GB TOTAL KNEE, Model Number: CDS984147K; 37) GB LOWER EXTREMITY, Model Number: CDS984156L; 38) LOWER EXTREMTIY, Model Number: CDS984254K; 39) LOWER EXTREMTIY, Model Number: CDS984254L; 40) LOWER EXTREMTIY, Model Number: CDS984254M; 41) LAMI MICRODISC-LF, Model Number: CDS984527I; 42) EXTREMITY CDS-LF, Model Number: CDS984607G; 43) EXTREMITY CDS-LF, Model Number: CDS984607I; 44) HANA/FRACTURE TABLE CDS-LF, Model Number: CDS984610C; 45) TOTAL JOINT CDS, Model Number: CDS985202B; 46) TOTAL JOINT CDS, Model Number: CDS985202C; 47) TRAUMA-LF, Model Number: CDS985279K; 48) ARTHROSCOPY CDS, Model Number: CDS985316A; 49) EXTREMITY CDS, Model Number: CDS985317C; 50) EXTREMITY CDS, Model Number: CDS985317D; 51) EXTREMITY CDS, Model Number: CDS985317F; 52) EXTREMITY CDS, Model Number: CDS985317G; 53) KNEE ARTHROSCOPY CDS, Model Number: CDS985425J; 54) TOTAL HIP PACK (CDSOPM834)-LF, Model Number: CDSOPM834W; 55) JOINT ASPIRATION KIT, Model Number: DYKM1161; 56) TOTAL KNEE SUPPLEMENT PACK NS, Model Number: DYKM2106; 57) KIT ORTHRO TOTAL HIP DR. NANDI, Model Number: DYKMBNDL137; 58) KIT NEUR FUS LUM POST CRANDALL, Model Number: DYKMBNDL138A; 59) KIT NEURO CRANIOTOMYANEURYSM R, Model Number: DYKMBNDL141; 60) KIT NEURO CRANIOTOMYANEURYSM R, Model Number: DYKMBNDL141A; 61) KIT NEURO CRANIOTOMYANEURYSM R, Model Number: DYKMBNDL141B; 62) KIT NEURO FUS LUM POST SANSUR, Model Number: DYKMBNDL142C; 63) KIT NEURO CRANIOTOMY ACOUSTIC, Model Number: DYKMBNDL143; 64) KIT NEURO CRANIOTOMY FOR TUMOR, Model Number: DYKMBNDL144; 65
Check your product
The notice names 54 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 10193489847260
- 40193489847261
- 10193489933758
- 40193489933759
- 10193489788655
- 40193489788656
- 10193489900972
- 40193489900973
- 10195327064495
- 40195327064496
- 10193489900989
- 40193489900980
- 10195327064501
- 40195327064502
Lot or batch code
Usually ink-stamped near the seam, the cap or the bottom of the pack.
- 21EBC900
- 22GBE442
- 22EBG548
- 22EBF695
- 22EBA274
- 22BBY060
- 21KBH163
- 21JBF936
- 21IBJ260
- 21GBI183
- 21FBS254
- 21BBL223
- 21LBO219
- 21KBQ650
- 21KBB625
- 21IBV695
- 21IBU197
- 21GBR895
- 23JBI054
- 23GBP400
Model or item number
On the rating label, the box or the instruction sheet.
- CDS940087AG
- CDS940087AI
- CDS980832Q
- CDS980832R
- CDS980832S
- CDS980833Q
- CDS980833R
- CDS980865V
- CDS980865W
- CDS980865X
- CDS980865Y
- CDS981454K
- CDS981454L
- CDS981502K
- CDS981505O
- CDS981505P
- CDS981505Q
- CDS981505R
- CDS981858F
- CDS982035W
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Sterility
Medline Industries, LP — The concern here is contamination of a product that bypasses the body’s normal defences, which is why the threshold for recalling it is low.
- Stop using the affected lot and check the lot number on your own carton or vial against the codes listed above.
- Do not use a vial or syringe with visible particles, cloudiness or discolouration, whether or not it is on the recall list.
- Call your prescriber or pharmacy today about a replacement from an unaffected lot — do not simply skip doses of a medicine you depend on.
- If you are a clinic or pharmacy, quarantine the stock and work it through your usual recall log.
Getting a replacement supply
Do not let a refund question interrupt your treatment. Deal with the supply first and the money second.
- Call the pharmacy before you run low. They can see which lot they dispensed to you and will usually swap it the same day.
- Insurers do not normally count a recall replacement against a refill limit. If one is refused, ask the pharmacist to call it through as a recall override.
- Over-the-counter products are refunded at the till like any other purchase; the recall notice on your phone is enough to explain the return.
- If the manufacturer runs a direct reimbursement form, it will be linked from the official notice on this page.
Company contact: Medline Industries, LP — 3 Lakes Dr, Northfield, IL, 60093-2753
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Medline Industries, LP medical devices recalled?
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products. Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 10193489847260, 40193489847261, 10193489933758; lot code 21EBC900, 22GBE442, 22EBG548; model number CDS940087AG, CDS940087AI, CDS980832Q. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Stop using the affected lot and check the lot number on your own carton or vial against the codes listed above.
Can I get my money back?
Call the pharmacy before you run low. They can see which lot they dispensed to you and will usually swap it the same day.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 11 days ago
Medline Industries, LP Rollator, Bariatric, Basic Steel, Blue recall: The front fork stem bolt/wheel assembly may disconnect from the rollator
The front fork stem bolt/wheel assembly may disconnect from the rollator frame tubing. If the front fork stem bolt/wheel assembly disconnects from…
-
Class II · 11 days ago
Medline Industries, LP Rollator, Bariatric, Basic Steel, Burgundy recall: The front fork stem bolt/wheel assembly may disconnect from the rollator
The front fork stem bolt/wheel assembly may disconnect from the rollator frame tubing. If the front fork stem bolt/wheel assembly disconnects from…
-
Class I · 18 days ago
Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories recall: Potential fire hazard involving the hand control pendant and associated
Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat…
-
Class I · 18 days ago
Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories recall: Potential fire hazard involving the hand control pendant and associated
Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat…
-
Class II · 25 days ago
Medline convenience kits containing Healthmark Instrument Channel recall: Kits contain Healthmark Instrument Channel Brushes
Kits contain Healthmark Instrument Channel Brushes, which were recalled by the manufacturer due to the potential for a sharp edge/burr on the end of…
-
Class II · 25 days ago
Medline Convenience Kits: 1) Anterior Spine Pack recall: The kits contain neurological sponges with the potential for unexpected
The kits contain neurological sponges with the potential for unexpected variability of endotoxin levels. Potentially elevated endotoxin levels may…