Class II U.S. Food and Drug Administration

Medline breathing circuits labeled as: 1) Medline Hudson RCI recall: Neonatal and infant heated wire breathing circuits may experience sporadic

Recall published last month

Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.

The hazard: Fire or burn hazard

The hazard here is thermal: the product can get hot enough to ignite what is around it, or to burn whoever is holding it.

What the notice says: Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.

Why it matters: The product can overheat, ignite or cause a burn. Where lithium batteries are involved the failure is fast and very hot, and water does not put it out quickly.

Symptoms and when to call a doctor

Treat a burn by cooling it under running water for about 20 minutes — not with ice, butter or ointment — and cover it loosely with a clean non-fluffy cloth. Get medical help for burns that are larger than the person’s palm, that are deep or white or charred, that involve the face, hands, feet or genitals, or any burn to a baby or young child. If someone has inhaled smoke and is coughing, hoarse or short of breath, call 911.

This page is general information, not medical advice. If you feel unwell, a clinician who can see you is the only person able to judge your case.

Recall details

Recall number
Z-2993-2026
Published
September 2, 2026
Risk level
Class II
Status
Ongoing
Issued by
U.S. Food and Drug Administration
Brand
Medline Industries, LP
Product type
Medical devices
Units affected
271406 units
Distribution
Nationwide
Made in
United States
Lot / code info
1) Medline Product Number/SKU HUD78001KIT (Teleflex Product Number/SKU 780-01KIT), UDI/DI 10197344010752 (each) 20197344010759 (case), All lot numbers; 2) Medline Product Number/SKU HUD78007KIT (Teleflex Product Number/SKU 780-07KIT), UDI/DI 10197344010738 (each) 20197344010735 (case), All lot numbers; 3) Medline Product Number/SKU HUD78008KIT (Teleflex Product Number/SKU 780-08KIT), UDI/DI 10197344010721 (each) 20197344010728 (case), All lot numbers; 4) Medline Product Number/SKU HUD78009KIT (Teleflex Product Number/SKU 790-09KIT), UDI/DI 10197344010714 (each) 20197344010711 (case), All lot numbers; 5) Medline Product Number/SKU HUD78017KIT (Teleflex Product Number/SKU 780-17KIT), UDI/DI 10197344010677 (each) 20197344010674 (case), All lot numbers; 6) Medline Product Number/SKU HUD78018KIT (Teleflex Product Number/SKU 780-18KIT), UDI/DI 10197344010905 (each) 20197344010902 (case), All lot numbers; 7) Medline Product Number/SKU HUD88001KIT (Teleflex Product Number/SKU 880-01KIT), UDI/DI 10197344010929 (each) 20197344010926 (case), All lot numbers; 8) Medline Product Number/SKU HUD88013KIT (Teleflex Product Number/SKU 880-13KIT), UDI/DI 10197344010493 (each) 20197344010490 (case), All lot numbers; 9) Medline Product Number/SKU HUD88014KIT (Teleflex Product Number/SKU 880-14KIT), UDI/DI 10197344010486 (each) 20197344010483 (case), All lot numbers; 10) Medline Product Number/SKU HUD88015KIT (Teleflex Product Number/SKU 880-15KIT), UDI/DI 10197344010479 (each) 20197344010476 (case), All lot numbers; 11) Medline Product Number/SKU HUD88017KIT (Teleflex Product Number/SKU 880-17KIT), UDI/DI 10197344010462 (each) 20197344010469 (case), All lot numbers;
UPC codes
10197344010752, 20197344010759, 10197344010738, 20197344010735, 10197344010721, 20197344010728, 10197344010714, 20197344010711, 10197344010677, 20197344010674, 10197344010905, 20197344010902, 10197344010929, 20197344010926, 10197344010493, 20197344010490, 10197344010486, 20197344010483, 10197344010479, 20197344010476

Products covered by this recall

Medline breathing circuits labeled as: 1) MEDLINE HUDSON RCI, CIRCUIT, NEO, SING LIMB, HTD,COLUMN, Medline Product Number/SKU HUD78001KIT (Teleflex Product Number/SKU 780-01KIT); 2) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78007KIT (Teleflex Product Number/SKU 780-07KIT); 3) MEDLINE HUDSON RCI, DBD-CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78008KIT (Teleflex Product Number/SKU 780-08KIT); 4) MEDLINE HUDSON RCI, DBD-CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78009KIT (Teleflex Product Number/SKU 790-09KIT); 5) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL LIMB, HTD,COLUMN, Medline Product Number/SKU HUD78017KIT (Teleflex Product Number/SKU 780-17KIT); 6) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL HTD, DUAL LIMB,COLUMN, Medline Product Number/SKU HUD78018KIT (Teleflex Product Number/SKU 780-18KIT); 7) MEDLINE HUDSON RCI, CIRCUIT,ISO-SHIELD SINGLE LIMB,INFANT, Medline Product Number/SKU HUD88001KIT (Teleflex Product Number/SKU 880-01KIT); 8) MEDLINE HUDSON RCI, CIRCUIT,ISO-SHIELD,INFANT,HTD,COLUMN,DRN, Medline Product Number/SKU HUD88013KIT (Teleflex Product Number/SKU 880-13KIT); 9) MEDLINE HUDSON RCI, ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88014KIT (Teleflex Product Number/SKU 880-14KIT); 10) MEDLINE HUDSON RCI,ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88015KIT (Teleflex Product Number/SKU 880-15KIT); 11) MEDLINE HUDSON RCI, ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88017KIT (Teleflex Product Number/SKU 880-17KIT).

Check your product

The notice names 40 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.

UPC / barcode

Read the twelve digits under the barcode on the package.

  • 10197344010752
  • 20197344010759
  • 10197344010738
  • 20197344010735
  • 10197344010721
  • 20197344010728
  • 10197344010714
  • 20197344010711
  • 10197344010677
  • 20197344010674
  • 10197344010905
  • 20197344010902
  • 10197344010929
  • 20197344010926
  • 10197344010493
  • 20197344010490
  • 10197344010486
  • 20197344010483
  • 10197344010479
  • 20197344010476

Model or item number

On the rating label, the box or the instruction sheet.

  • SKU HUD78001KIT
  • SKU 780-01KIT
  • SKU HUD78007KIT
  • SKU 780-07KIT
  • SKU HUD78008KIT
  • SKU 780-08KIT
  • SKU HUD78009KIT
  • SKU 790-09KIT
  • SKU HUD78017KIT
  • SKU 780-17KIT
  • SKU HUD78018KIT
  • SKU 780-18KIT
  • SKU HUD88001KIT
  • SKU 880-01KIT
  • SKU HUD88013KIT
  • SKU 880-13KIT
  • SKU HUD88014KIT
  • SKU 880-14KIT
  • SKU HUD88015KIT
  • SKU 880-15KIT

Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.

What to do about Fire

Medline Industries, LP — The hazard here is thermal: the product can get hot enough to ignite what is around it, or to burn whoever is holding it.

  1. Stop using it and unplug it from the wall. If it runs on a battery pack, stop charging it.
  2. Move it off anything that burns — a bed, a sofa, a carpet, a pile of paper — and away from exits.
  3. Do not leave it charging overnight or while the house is empty while you wait for the remedy.
  4. If it has already overheated, scorched or swelled, do not keep using it and do not try to repair it.

If the claim stalls

Most recall claims are paid without argument. The ones that stall usually stall for a reason you can fix.

  • Check you used the recall channel rather than ordinary customer support — the two queues rarely talk to each other.
  • Paid by card? A chargeback for goods that are unfit for sale is a legitimate fallback once the company has had a fair chance.
  • Send the photos they asked for in the format they asked for. A missing date code is the single commonest reason a claim sits unprocessed.
  • Give it ten working days, then chase in writing and quote your claim reference.

Company contact: Medline Industries, LP — 3 Lakes Dr, Northfield, IL, 60093-2753

The official notice

This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.

Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls

Questions people ask about this recall

Why was Medline Industries, LP medical devices recalled?

Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress. Neonatal and infant heated wire breathing circuits may…

How do I know if I have the recalled product?

Match the codes printed on your package against the notice: UPC 10197344010752, 20197344010759, 10197344010738; model number SKU HUD78001KIT, SKU 780-01KIT, SKU HUD78007KIT. If the code on your package is not on that list, your unit is outside the recall.

What should I do with it right now?

Stop using it and unplug it from the wall. If it runs on a battery pack, stop charging it.

Can I get my money back?

Check you used the recall channel rather than ordinary customer support — the two queues rarely talk to each other.

Where was the recalled product sold?

It was distributed nationwide.