Class I U.S. Food and Drug Administration

Medline beds labeled as: Full Electric Low Basic Homecare Bed recall: Medline Industries has identified a potential hazard involving the hand

Recall published 9 months ago

Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.

The hazard: Fire or burn hazard

The hazard here is thermal: the product can get hot enough to ignite what is around it, or to burn whoever is holding it.

What the notice says: Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.

Why it matters: The product can overheat, ignite or cause a burn. Where lithium batteries are involved the failure is fast and very hot, and water does not put it out quickly.

Symptoms and when to call a doctor

Treat a burn by cooling it under running water for about 20 minutes — not with ice, butter or ointment — and cover it loosely with a clean non-fluffy cloth. Get medical help for burns that are larger than the person’s palm, that are deep or white or charred, that involve the face, hands, feet or genitals, or any burn to a baby or young child. If someone has inhaled smoke and is coughing, hoarse or short of breath, call 911.

This page is general information, not medical advice. If you feel unwell, a clinician who can see you is the only person able to judge your case.

Recall details

Recall number
Z-1014-2026
Published
January 28, 2026
Risk level
Class I
Status
Ongoing
Issued by
U.S. Food and Drug Administration
Brand
Medline Industries, LP
Product type
Medical devices
Units affected
22565 units
Distribution
California, Guam, Pennsylvania, U.S. Virgin Islands
Made in
United States
Lot / code info
UDI/DI 10080196546992, All lots
UPC codes
10080196546992

Products covered by this recall

MEDLINE beds labeled as: Full Electric Low Basic Homecare Bed, REF MDR107003ELO

Check your product

The notice names 5 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.

UPC / barcode

Read the twelve digits under the barcode on the package.

  • 10080196546992

Lot or batch code

Usually ink-stamped near the seam, the cap or the bottom of the pack.

  • s — 22565 units

Model or item number

On the rating label, the box or the instruction sheet.

  • MDR107003ELO — UDI
  • DI 10080196546992
  • All lots — 22

Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.

What to do about Fire

Medline Industries, LP — The hazard here is thermal: the product can get hot enough to ignite what is around it, or to burn whoever is holding it.

  1. Stop using it and unplug it from the wall. If it runs on a battery pack, stop charging it.
  2. Move it off anything that burns — a bed, a sofa, a carpet, a pile of paper — and away from exits.
  3. Do not leave it charging overnight or while the house is empty while you wait for the remedy.
  4. If it has already overheated, scorched or swelled, do not keep using it and do not try to repair it.

If the claim stalls

Most recall claims are paid without argument. The ones that stall usually stall for a reason you can fix.

  • Check you used the recall channel rather than ordinary customer support — the two queues rarely talk to each other.
  • Paid by card? A chargeback for goods that are unfit for sale is a legitimate fallback once the company has had a fair chance.
  • Send the photos they asked for in the format they asked for. A missing date code is the single commonest reason a claim sits unprocessed.
  • Give it ten working days, then chase in writing and quote your claim reference.

Company contact: Medline Industries, LP — 3 Lakes Dr, Northfield, IL, 60093-2753

The official notice

This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.

Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls

Questions people ask about this recall

Why was Medline Industries, LP medical devices recalled?

Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire. Medline Industries has identified a potential hazard involving the hand…

How do I know if I have the recalled product?

Match the codes printed on your package against the notice: UPC 10080196546992; lot code s — 22565 units; model number MDR107003ELO — UDI, DI 10080196546992, All lots — 22. If the code on your package is not on that list, your unit is outside the recall.

What should I do with it right now?

Stop using it and unplug it from the wall. If it runs on a battery pack, stop charging it.

Can I get my money back?

Check you used the recall channel rather than ordinary customer support — the two queues rarely talk to each other.

Which states were affected?

The notice lists California, Guam, Pennsylvania, U.S. Virgin Islands.

Related recalls