Interventional Fluoroscopic X-Ray System - Philips Allura Xper Series recall: Potential that a component (air baffle) may be missing in some Nehalem
Potential that a component (air baffle) may be missing in some Nehalem FlexVision-2 PCs that are part of the Allura Xper series.
The hazard: Device safety concern
The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
What the notice says: Potential that a component (air baffle) may be missing in some Nehalem FlexVision-2 PCs that are part of the Allura Xper series.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-2504-2025
- Published
- September 10, 2025
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Philips Medical Systems Nederland B.v.
- Product type
- Medical devices
- Units affected
- 192 untis
- Distribution
- Nationwide
- Made in
- Netherlands
- Lot / code info
- 1. Model Number/System Code (UDI); Serial Numbers: 722003 (No UDI); 292, 314, 1077, 464. 2. Model Number/System Code (UDI); Serial Numbers: 722005 (No UDI); 54. 3. Model Number/System Code (UDI); Serial Numbers: 722006 (No UDI); 2126, 2200, 1560, 1921, 1724, 1219. 4. Model Number/System Code (UDI); Serial Numbers: 722008 (No UDI); 516, 142, 434, 418, 480. 5. Model Number/System Code (UDI); Serial Numbers: 722010 (00884838059030); 33, 968, 763, 215, 1425, 1126, 818, 1028, 1043, 608, 907, 50, 1390, 706, 900, 1439, 1152, 165, 62, 1031, 1088, 741, 840, 413. 6. Model Number/System Code (UDI); Serial Numbers: 722011 (00884838059047); 308, 344, 243, 266, 231, 286, 369, 378, 259, 333. 7. Model Number/System Code (UDI); Serial Numbers: 722012 (00884838059054); 2898, 2896, 1594, 1249, 2057, 2135, 2857, 792, 2841, 1334, 1550, 2868, 2912, 1094, 2282, 1681, 1332, 2032, 2076, 176, 835, 626, 1320, 1439, 1667. 8. Model Number/System Code (UDI); Serial Numbers: 722013 (00884838059061); 115, 575, 272, 632, 378, 608, 412, 353, 595, 355, 403. 9. Model Number/System Code (UDI); Serial Numbers: 722023 (00884838059085); 128, 83, 86, 134, 87. 10. Model Number/System Code (UDI); Serial Numbers: 722026 (00884838054189); 233, 561, 942, 368, 826, 1115, 539, 159, 598, 365, 528, 603, 263, 792, 783, 679, 680, 406, 388, 302, 659, 736, 1022, 22, 506, 120, 526. 11. Model Number/System Code (UDI); Serial Numbers: 722027 (00884838054196); 106, 127, 78, 204, 113, 82, 66, 119. 12. Model Number/System Code (UDI); Serial Numbers: 722028 (00884838054202); 668, 265, 1154, 180, 880, 274, 172, 1111, 806, 451, 236, 887, 874, 1266, 1499, 1514, 1457, 908, 1292, 247, 944, 951, 1390, 1662, 415, 105, 1675, 348, 1562, 1029, 51, 941, 403, 427, 659, 935, 1225, 1165, 1621, 725, 723, 1559, 753, 883. 13. Model Number/System Code (UDI); Serial Numbers: 722029 (00884838054219); 69, 58. 14. Model Number/System Code (UDI); Serial Numbers: 722035 (00884838054240); 187, 125, 40, 108, 163, 26, 75, 173, 190, 103. 15. Model Number/System Code (UDI); Serial Numbers: 722038 (00884838054226); 63. 16. Model Number/System Code (UDI); Serial Numbers: 722058 (00884838059115); 238, 161, 84, 39, 133, 244, 193, 187, 102.
- UPC codes
- 00884838059030, 00884838059047, 00884838059054, 00884838059061, 00884838059085, 00884838054189, 00884838054196, 00884838054202, 00884838054219, 00884838054240, 00884838054226, 00884838059115
Products covered by this recall
Interventional Fluoroscopic X-Ray System - Philips Allura Xper Series. Labeled as the following with corresponding model numbers: 1. Allura Xper FD10, Model Number 722003. 2. Allura Xper FD10/10, Model Number 722005. 3. Allura Xper FD20, Model Number 722006. 4. Allura Xper FD20 Biplane, Model Number 722008. 5. Allura Xper FD10, Model Number 722010. 6. Allura Xper FD10/10, Model Number 722011. 7. Allura Xper FD20, Model Number 722012. 8. Allura Xper FD20 Biplane, Model Number 722013. 9. Allura Xper FD20 OR Table, Model Number 722023. 10. Allura Xper FD10, Model Number 722026. 11. Allura Xper FD10/10, Model Number 722027. 12. Allura Xper FD20, Model Number 722028. 13. Allura Xper FD20/10, Model Number 722029. 14. Allura Xper FD20 OR Table, Model Number 722035. 15. Allura Xper FD20/20, Model Number 722038. 16. Allura Xper FD20/15, Model Number 722058. 510(k) Numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859.
Check your product
The notice names 32 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 00884838059030
- 00884838059047
- 00884838059054
- 00884838059061
- 00884838059085
- 00884838054189
- 00884838054196
- 00884838054202
- 00884838054219
- 00884838054240
- 00884838054226
- 00884838059115
Model or item number
On the rating label, the box or the instruction sheet.
- 1. Allura Xper FD10
- 10
- 20
- 722058. 510
Serial number
On the rating plate, often next to the date of manufacture.
- 722003
- 722005
- 722006
- 722008
- 722010
- 722011
- 722012
- 722013
- 722023
- 722026
- 722027
- 722028
- 722029
- 722035
- 722038
- 722058
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Philips Medical Systems Nederland B.v. — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
Who pays for a device recall
The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.
- If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
- Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
- If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
- Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.
Company contact: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Veenpluis 4-6, Best
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Philips Medical Systems Nederland B.v. medical devices recalled?
Potential that a component (air baffle) may be missing in some Nehalem FlexVision-2 PCs that are part of the Allura Xper series. Potential that a component (air baffle) may be missing in some Nehalem FlexVision-2 PCs that are part of the Allura Xper series.
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 00884838059030, 00884838059047, 00884838059054; model number 1. Allura Xper FD10, 10, 20. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 2 months ago
Philips Medical Systems Nederland B.v. Allura Xper FD. Model Numbers: 722026 recall: Firmware issue potentially leading to loss of geometry movements
Firmware issue potentially leading to loss of geometry movements.
-
Class II · 2 months ago
Philips Azurion, Model Numbers: 722229, 722230, 722231, 722232 recall: Two software related issues: 1) Potential image quality degradation
Two software related issues: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate…
-
Class II · 2 months ago
Philips Azurion Rystem, Model Numbers: 722063, 722064, 722067 recall: Two software related issues affecting Azurion R2.2.10 Systems: 1) Potential
Two software related issues affecting Azurion R2.2.10 Systems: 1) Potential image quality degradation that may lead to delay of therapy, procedural…
-
Class II · 2 months ago
Philips Allura, Cardio Vascular-Allura Centron recall: Potential Loss of imaging functionality, loss of motorized movement
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
-
Class II · 2 months ago
Philips Azurion under the following System Descriptions recall: Potential Loss of imaging functionality, loss of motorized movement
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
-
Class II · 2 months ago
Philips Allura under the following System Descriptions recall: Potential Loss of imaging functionality, loss of motorized movement
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.