Class II U.S. Food and Drug Administration

Integra LifeSciences Corp. TruDi¿ Navigation System - Multi Instrument Adapter recall: It was determined that when the navigation system is configured with specific

Recall published 12 months ago

It was determined that when the navigation system is configured with specific revision combinations of the Multi Instrument Adapter together with the Patient Tracker the system may not meet its specified accuracy for visual verification of device location within the patient s anatomy.

The hazard: Device safety concern

The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.

What the notice says: It was determined that when the navigation system is configured with specific revision combinations of the Multi Instrument Adapter together with the Patient Tracker the system may not meet its specified accuracy for visual verification of device location within the patient s anatomy.

Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.

Recall details

Recall number
Z-0420-2026
Published
November 12, 2025
Risk level
Class II
Status
Ongoing
Issued by
U.S. Food and Drug Administration
Brand
Integra LifeSciences Corp.
Product type
Medical devices
Units affected
1198 units
Distribution
Nationwide
Made in
United States
Lot / code info
Manufacturer's Product Number (Catalog Number): EM-2025-000F Rev 00 or Rev 01; UDI-DI: 10846835021103;
UPC codes
10846835021103

Products covered by this recall

TruDi¿ Navigation System - Multi Instrument Adapter; - Manufacturer's Product Number (Catalog Number): EM-2025-000F Rev 00 or Rev 01;

Check your product

The notice names 1 code you can read straight off the package. Type the one printed on yours and the list below narrows to it.

UPC / barcode

Read the twelve digits under the barcode on the package.

  • 10846835021103

Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.

What to do about Device

Integra LifeSciences Corp. — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.

  1. Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
  2. If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
  3. Ask the supplier or clinic that provided it what the remedy is in your case.
  4. If you are a facility, quarantine affected stock and log the correction.

Who pays for a device recall

The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.

  • If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
  • Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
  • If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
  • Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.

Company contact: Integra LifeSciences Corp. (NeuroSciences) — 1100 Campus Rd, Princeton, NJ, 08540-6650

The official notice

This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.

Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls

Questions people ask about this recall

Why was Integra LifeSciences Corp. medical devices recalled?

It was determined that when the navigation system is configured with specific revision combinations of the Multi Instrument Adapter together with the Patient Tracker the system may not meet its specified accuracy for visual verification of device location within the patient s anatomy. It was determined that when the navigation system is configured with specific revision combinations of the Multi Instrument Adapter together with the Patient Tracker the system may not meet its specified accuracy for visual verification of device location within the patient s anatomy.

How do I know if I have the recalled product?

Match the codes printed on your package against the notice: UPC 10846835021103. If the code on your package is not on that list, your unit is outside the recall.

What should I do with it right now?

Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.

Can I get my money back?

If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.

Where was the recalled product sold?

It was distributed nationwide.