Instrumentation Laboratory HemoslL AcuStar HIT-lgG recall: Potential for invalid calibration curve or negative mean bias of up to 47%
Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.
The hazard: Device safety concern
This is a device enforcement record. Check whether the remedy is a correction, a software update or a removal before changing anything.
What the notice says: Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-0001-2027
- Published
- October 21, 2026
- Risk level
- Class I
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Instrumentation Laboratory
- Product type
- Medical devices
- Units affected
- 434 units
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- Part Number: 0009800127. UDI-DI: 08426951126545. Lot Number (Expiration Date): B40005 (01/16/2027), B38062 (11/01/2026).
Products covered by this recall
HemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009800127. in vitro diagnostic
Check your product
The notice names 2 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
Model or item number
On the rating label, the box or the instruction sheet.
- 0009800127
- 0009800127. UDI-DI
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Instrumentation Laboratory — This is a device enforcement record. Check whether the remedy is a correction, a software update or a removal before changing anything.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- Ask the supplier or clinic that provided it what the remedy is in your case.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
Replacements, repairs and reimbursement
Device remedies run through the supply chain, so the route depends on who sold you the unit.
- Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
- If the device was billed to insurance or Medicare, the replacement is usually billed the same way; you should not face a second deductible for a recall.
- Keep every piece of correspondence. If a recalled device caused harm, that paper trail is what a claim rests on.
- Consumer-purchased devices such as monitors and test kits are normally replaced free, with return shipping paid.
Company contact: Instrumentation Laboratory — 180 Hartwell Rd, Bedford, MA, 01730-2443
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Instrumentation Laboratory medical devices recalled?
Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples. Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: model number 0009800127, 0009800127. UDI-DI. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class I · 7 days ago
Instrumentation Laboratory HemoslL AcuStar HIT-lgG recall: Potential for invalid calibration curve or negative mean bias of up to 47%
Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.
-
Class II · 3 months ago
Instrumentation Laboratory GEM Premier 5000 PAK Part Numbers: 00055360010 00055360011 recall: The active ingredient concentration in the lysing bag solution of specific
The active ingredient concentration in the lysing bag solution of specific GEM Premier 5000 lots did not meet established manufacturing…
-
Class II · 3 months ago
Instrumentation Laboratory GEM Premier 7000 with iQM3 Part Numbers: 00077407508 recall: The active ingredient concentration in the lysing bag solution of specific
The active ingredient concentration in the lysing bag solution of specific GEM Premier 7000 with iQM3 PAK lots did not meet established manufacturing…
-
Class II · 3 months ago
Instrumentation Laboratory Brand Name: GEM Premier 5000 Product Name: GEM Premier recall: Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000
Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect…
-
Class II · 3 months ago
Instrumentation Laboratory Brand Name: GEM Premier 5000 Product Name: GEM Premier recall: Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000
Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect…
-
Class II · 3 months ago
Instrumentation Laboratory Brand Name: GEM Premier 5000 Product Name: GEM Premier recall: Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000
Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect…