Class II U.S. Food and Drug Administration

Howmedica Osteonics Corp. Trident 0¿ Poly Insert 28mm ID 620-00-28D Orthopaedic recall: Affected devices were manufactured with Ultra High Molecular Weight

Recall published last month

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

The hazard: Device safety concern

A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.

What the notice says: Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.

Recall details

Recall number
Z-3204-2026
Published
September 23, 2026
Risk level
Class II
Status
Ongoing
Issued by
U.S. Food and Drug Administration
Brand
Howmedica Osteonics Corp.
Product type
Medical devices
Units affected
12,156 US
Distribution
Nationwide
Made in
United States
Lot / code info
620-00-28D UDI-DI 07613327313581 Lots MHANP2 MHD1N0
UPC codes
07613327313581

Products covered by this recall

TRIDENT 0¿ POLY INSERT 28mm ID 620-00-28D Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Check your product

The notice names 3 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.

UPC / barcode

Read the twelve digits under the barcode on the package.

  • 07613327313581

Lot or batch code

Usually ink-stamped near the seam, the cap or the bottom of the pack.

  • MHANP2 MHD1N0 — 12
  • 156 US

Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.

What to do about Device

Howmedica Osteonics Corp. — A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.

  1. Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
  2. If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
  3. Ask the supplier or clinic that provided it what the remedy is in your case.
  4. If you are a facility, quarantine affected stock and log the correction.

Claiming on a recalled device

Device recalls are tracked by serial and lot number, so your claim moves faster if you lead with those.

  • Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.
  • Ask explicitly whether your unit is in the affected range; many recalls cover only part of a model line.
  • If you need the device daily, ask what interim arrangement they offer while the remedy is carried out.
  • Log the date, the name and the reference number of every call. Device recalls can run for months.

Company contact: Howmedica Osteonics Corp. — 325 Corporate Dr, Mahwah, NJ, 07430-2006

The official notice

This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.

Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls

Questions people ask about this recall

Why was Howmedica Osteonics Corp. medical devices recalled?

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation. Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of…

How do I know if I have the recalled product?

Match the codes printed on your package against the notice: UPC 07613327313581; lot code MHANP2 MHD1N0 — 12, 156 US. If the code on your package is not on that list, your unit is outside the recall.

What should I do with it right now?

Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.

Can I get my money back?

Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.

Where was the recalled product sold?

It was distributed nationwide.

Related recalls