Class II U.S. Food and Drug Administration

Hologic, Inc 3Dimensions Mammography System recall: Complaints have been received of systems developing loose, missing

Recall published 5 months ago

Complaints have been received of systems developing loose, missing, or broken internal bolts over time.

The hazard: Device safety concern

The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.

What the notice says: Complaints have been received of systems developing loose, missing, or broken internal bolts over time.

Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.

Recall details

Recall number
Z-2168-2026
Published
May 20, 2026
Risk level
Class II
Status
Ongoing
Issued by
U.S. Food and Drug Administration
Brand
Hologic, Inc
Product type
Medical devices
Units affected
14837 units
Distribution
Nationwide
Made in
United States
Lot / code info
"[Model No.; UDI]: [3DM-SYS-INTL2D;15420045510609], [3DM-SYS-INTL2D-MOB;15420045510593], [3DM-SYS-INTL2D-NS;15420045510586], [3DM-SYS-INTL3D;15420045510579], [3DM-SYS-INTL3D-MOB;15420045510562], [3DM-SYS-INTL3D-NS;15420045510555], [3DM-SYS-STD;15420045510524], [3DM-SYS-STD-MOB;15420045510531], [3DM-SYS-STD-NS;15420045510548]; All Serial No.
UPC codes
15420045510609, 15420045510593, 15420045510586, 15420045510579, 15420045510562, 15420045510555, 15420045510524, 15420045510531, 15420045510548

Products covered by this recall

3DIMENSIONS MAMMOGRAPHY SYSTEM; Model Numbers: 3DM-SYS-INTL2D, 3DM-SYS-INTL2D-MOB, 3DM-SYS-INTL2D-NS, 3DM-SYS-INTL3D, 3DM-SYS-INTL3D-MOB, 3DM-SYS-INTL3D-NS, 3DM-SYS-STD, 3DM-SYS-STD-MOB, 3DM-SYS-STD-NS.

Check your product

The notice names 12 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.

UPC / barcode

Read the twelve digits under the barcode on the package.

  • 15420045510609
  • 15420045510593
  • 15420045510586
  • 15420045510579
  • 15420045510562
  • 15420045510555
  • 15420045510524
  • 15420045510531
  • 15420045510548

Model or item number

On the rating label, the box or the instruction sheet.

  • 3DM-SYS-INTL2D
  • 3DM-SYS-INTL2D-MOB

Serial number

On the rating plate, often next to the date of manufacture.

  • No. — 14837 units

Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.

What to do about Device

Hologic, Inc — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.

  1. Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
  2. If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
  3. Ask the supplier or clinic that provided it what the remedy is in your case.
  4. If you are a facility, quarantine affected stock and log the correction.

Who pays for a device recall

The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.

  • If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
  • Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
  • If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
  • Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.

Company contact: Hologic, Inc — 250 Campus Dr, Marlborough, MA, 01752-3020

The official notice

This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.

Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls

Questions people ask about this recall

Why was Hologic, Inc medical devices recalled?

Complaints have been received of systems developing loose, missing, or broken internal bolts over time. Complaints have been received of systems developing loose, missing, or broken internal bolts over time.

How do I know if I have the recalled product?

Match the codes printed on your package against the notice: UPC 15420045510609, 15420045510593, 15420045510586; model number 3DM-SYS-INTL2D, 3DM-SYS-INTL2D-MOB. If the code on your package is not on that list, your unit is outside the recall.

What should I do with it right now?

Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.

Can I get my money back?

If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.

Where was the recalled product sold?

It was distributed nationwide.

Related recalls