GE Medical Systems, LLC Revolution Ascend, computed tomography recall: GE HealthCare has become aware of a potential security vulnerability
GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
The hazard: Device safety concern
A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.
What the notice says: GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-2149-2026
- Published
- May 20, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- GE Medical Systems, LLC
- Product type
- Medical devices
- Units affected
- 25 units
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- UDI-DI: 00840682146173; Serial Numbers: AS10G2200136YC AS10G2200101YC AS10G2200068YC AS10G2200035YC AS10G2300093YC AS10G2400094YC AS10G2400079YC AS10G2100008YC AS10G2500002YC CBDWG2400019HM CBDWG2300081HM AS10G2400056YC AS10G2400081YC AS10G2400087YC CBDWG2400117HM CBDWG2400085HM CBDWG2500003HM CBDWG2400140HM CBDWG2400126HM AS10G2200085YC AS10G2300057YC CBDWG2500011HM CBDWG2300056HM CBDWG2300041HM AS10G2300041YC
- UPC codes
- 00840682146173
Products covered by this recall
Revolution Ascend, computed tomography, Model Numbers 6969000-100 and 6969000-300
Check your product
The notice names 4 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 00840682146173
Model or item number
On the rating label, the box or the instruction sheet.
- 6969000-100
- 6969000-300 — UDI-DI
Serial number
On the rating plate, often next to the date of manufacture.
- AS10G2200136YC AS10G2200101YC AS10G220006
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
GE Medical Systems, LLC — A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- Ask the supplier or clinic that provided it what the remedy is in your case.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
Claiming on a recalled device
Device recalls are tracked by serial and lot number, so your claim moves faster if you lead with those.
- Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.
- Ask explicitly whether your unit is in the affected range; many recalls cover only part of a model line.
- If you need the device daily, ask what interim arrangement they offer while the remedy is carried out.
- Log the date, the name and the reference number of every call. Device recalls can run for months.
Company contact: GE Medical Systems, LLC — 3000 N Grandview Blvd, Waukesha, WI, 53188-1615
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was GE Medical Systems, LLC medical devices recalled?
GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems. GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 00840682146173; model number 6969000-100, 6969000-300 — UDI-DI. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 3 months ago
GE Medical Systems, LLC Optima XR200amx, Mobile Digital-Ready Radiographic System recall: GE HealthCare has become aware that in certain Optima XR200/220/240amx
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of…
-
Class II · 3 months ago
GE Medical Systems, LLC Optima XR220amx, Mobile Digital-Ready Radiographic System recall: GE HealthCare has become aware that in certain Optima XR200/220/240amx
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of…
-
Class II · 3 months ago
GE Medical Systems, LLC Optima XR240amx, Mobile Digital Radiographic System recall: GE HealthCare has become aware that in certain Optima XR200/220/240amx
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of…
-
Class II · 4 months ago
GE Medical Systems, LLC Giraffe OmniBed Field Replaceable Unit recall: GE HealthCare has become aware of a potential issue involving the canopy
GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain…
-
Class II · 4 months ago
GE Medical Systems, LLC Giraffe OmniBed, a combination of an infant incubator recall: GE HealthCare has become aware of a potential issue involving the canopy
GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain…
-
Class II · 4 months ago
GE Medical Systems, LLC Giraffe OmniBed Carestations and Giraffe OmniBed Carestation recall: GE HealthCare has become aware of a potential issue involving the canopy
GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain…