Class II U.S. Food and Drug Administration

GE Medical Systems, LLC MAC VU360 Acquisition Trunk Cable and Module Holder recall: If a user incorrectly places the Acquisition Module into the Acquisition

Recall published 12 months ago

If a user incorrectly places the Acquisition Module into the Acquisition Module Holder, it can result in excessive bending and wear of the Acquisition Trunk Cable. This is visible as a damaged sheath or exposed wires. If a damaged Acquisition Trunk Cable is not recognized and removed from service, the cable can overheat with continued use.

The hazard: Fire or burn hazard

A fire or burn risk is what prompted this recall. Most of these incidents happen while the product is charging or left running unattended.

What the notice says: If a user incorrectly places the Acquisition Module into the Acquisition Module Holder, it can result in excessive bending and wear of the Acquisition Trunk Cable. This is visible as a damaged sheath or exposed wires. If a damaged Acquisition Trunk Cable is not recognized and removed from service, the cable can overheat with continued use.

Why it matters: The product can overheat, ignite or cause a burn. Where lithium batteries are involved the failure is fast and very hot, and water does not put it out quickly.

Symptoms and when to call a doctor

Treat a burn by cooling it under running water for about 20 minutes — not with ice, butter or ointment — and cover it loosely with a clean non-fluffy cloth. Get medical help for burns that are larger than the person’s palm, that are deep or white or charred, that involve the face, hands, feet or genitals, or any burn to a baby or young child. If someone has inhaled smoke and is coughing, hoarse or short of breath, call 911.

This page is general information, not medical advice. If you feel unwell, a clinician who can see you is the only person able to judge your case.

Recall details

Recall number
Z-0231-2026
Published
October 29, 2025
Risk level
Class II
Status
Ongoing
Issued by
U.S. Food and Drug Administration
Brand
GE Medical Systems, LLC
Product type
Medical devices
Units affected
29,225 units
Distribution
Nationwide
Made in
United States
Lot / code info
UDI-DI: 00840682125499; Serial Numbers: VU221240028SA VU221250168SA VU222120101SA VU222120114SA VU222120120SA VU222400056SA VU222400070SA VU222480182SA VU421010035SA VU420530073SA VU420530105SA VU420530108SA VU420530109SA VU420530110SA VU421010007SA VU421010008SA VU421010009SA VU421010010SA VU421010012SA VU421010021SA VU421010029SA VU421010040SA VU421010042SA VU422410219SA VU422410220SA VU220500166SA VU220500167SA VU222250167SA VU222250175SA VU423290038SA VU423290039SA VU423290063SA VU423290068SA VU221500205SA VU221500215SA VU221500216SA VU221500218SA VU221300012SA VU221320090SA VU221330009SA VU221330012SA VU221330013SA VU221120005SA VU221490009SA VU221490015SA VU221490031SA VU221490033SA VU221490038SA VU221490043SA VU221490113SA VU421250084SA VU420290004SA VU420290052SA VU221350037SA VU421340025SA VU421340038SA VU421340117SA VU421340120SA VU421340127SA VU421340131SA VU421340139SA VU421340142SA VU421340163SA VU421350169SA VU421350174SA VU421350179SA VU421350201SA VU421350206SA VU421350207SA VU421350209SA VU421350210SA VU421350211SA VU421350213SA VU421350214SA VU421350216SA VU421350223SA VU421350225SA VU421350227SA VU421350228SA VU421350231SA VU421350233SA VU421350239SA VU421350248SA VU421350249SA VU421350251SA VU421350252SA VU421350253SA VU421350267SA VU421350268SA VU421350269SA VU421350271SA VU421350272SA VU421350273SA VU421350275SA VU421350279SA VU421350280SA VU421350282SA VU421350287SA VU421500058SA VU221010086SA VU221010089SA VU221020043SA VU221020044SA VU221020046SA VU221020047SA VU221020048SA VU221020051SA VU221020052SA VU221020054SA VU221020055SA VU221020056SA VU220100042SA VU220100050SA VU220100051SA VU220100055SA VU220100069SA VU220100073SA VU220100074SA VU220100076SA VU220100077SA VU220100079SA VU220100080SA VU220100082SA VU222390186SA VU221490003SA VU221490006SA VU221490010SA VU221490011SA VU221490012SA VU221490013SA VU221490014SA VU221490016SA VU221490021SA VU221490024SA VU221490029SA VU221490044SA VU221110086SA VU221040087SA VU222010022SA VU222010074SA VU222010078SA VU222010084SA VU222010096SA VU222010097SA VU220510054SA VU220510083SA VU220510084SA VU220310048SA VU220310071SA VU222060140SA VU222070017SA VU222070028SA VU222070081SA VU222070082SA VU222070087SA VU222070124SA VU222070131SA VU222070133SA VU222070135SA VU222070137SA VU222070140SA VU222070142SA VU222070143SA VU222070145SA VU222120095SA VU222340060SA VU222350205SA VU223240164SA VU223240206SA VU420290059SA VU421350234SA VU221490032SA VU223220131SA VU421380150SA VU421380154SA VU422480255SA VU422510026SA VU422510055SA VU422510063SA VU221380182SA VU420300098SA VU420300099SA VU420300106SA VU422010046SA VU422010051SA VU422010064SA VU422010111SA VU422020010SA VU422020016SA VU423050025SA VU423050026SA VU423050028SA VU423050030SA VU423050044SA VU423050045SA VU423050046SA VU423050047SA VU423050048SA VU423050079SA VU423050081SA VU423050082SA VU423050083SA VU423050085SA VU423050089SA VU423050090SA VU423050091SA VU422350040SA VU422350042SA VU422350043SA VU422350045SA VU423050078SA VU223110237SA VU223110246SA VU222230077SA VU222240008SA VU222240009SA VU223170029SA VU223170064SA VU223170066SA VU223170067SA VU223110337SA VU220440024SA VU220440038SA VU221130056SA VU223180099SA VU420300102SA VU420300120SA VU420300122SA VU422120361SA VU422120364SA VU422120367SA VU422120372SA VU422120376SA VU422120378SA VU422120380SA VU220020035SA VU223110352SA VU223110358SA VU221290085SA VU221290092SA VU221290095SA VU221290096SA VU221290102SA VU222020009SA VU220420030SA VU220420047SA VU220420075SA VU220420091SA VU220420095SA VU220430043SA VU220430051SA VU220430053SA VU223130175SA VU222260377SA VU222260380SA VU221260142SA VU221270082SA VU219500036SA VU219500046SA VU220110020SA VU220120003SA VU422130033SA VU421340033SA VU421340063SA VU421340068SA VU421340082SA VU421340083SA VU421340087SA VU421340099SA VU421340101SA VU421340111SA VU421340112SA VU422120362SA VU422120365SA VU422120368SA VU422120374SA VU422350022SA VU422350053SA VU422350057SA VU422350061SA VU422350094SA VU422350095SA VU422360043…
UPC codes
00840682125499

Products covered by this recall

MAC VU360 Acquisition Trunk Cable and Module Holder, high performance, multichannel resting electrocardiograph, Model Number 2030360-001, Product Codes VU2 and VU4

Check your product

The notice names 4 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.

UPC / barcode

Read the twelve digits under the barcode on the package.

  • 00840682125499

Model or item number

On the rating label, the box or the instruction sheet.

  • 2030360-001
  • VU4 — UDI-DI

Serial number

On the rating plate, often next to the date of manufacture.

  • VU221240028SA VU221250168SA VU222120101SA

Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.

What to do about Fire

GE Medical Systems, LLC — A fire or burn risk is what prompted this recall. Most of these incidents happen while the product is charging or left running unattended.

  1. Stop using it and unplug it from the wall. If it runs on a battery pack, stop charging it.
  2. Check that your smoke alarms work while you are thinking about it, and follow the remedy on this page.
  3. Move it off anything that burns — a bed, a sofa, a carpet, a pile of paper — and away from exits.
  4. Do not leave it charging overnight or while the house is empty while you wait for the remedy.

How the remedy works

CPSC recalls specify one of three remedies, and which one you get is not negotiable — but how smoothly it goes often is.

  • A refund means the purchase price back, sometimes prorated by age for durable goods. Ask how they calculated it if the figure looks low.
  • Several recalls now ask you to destroy the item and send a photo as proof. Register the claim before you do that, never after.
  • Return shipping is the company’s cost. If you are asked to pay it, push back.
  • A repair kit is shipped free and is designed to be fitted at home; if you cannot fit it, ask whether they will arrange installation.

Company contact: GE Medical Systems, LLC — 3000 N Grandview Blvd, Waukesha, WI, 53188-1615

The official notice

This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.

Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls

Questions people ask about this recall

Why was GE Medical Systems, LLC medical devices recalled?

If a user incorrectly places the Acquisition Module into the Acquisition Module Holder, it can result in excessive bending and wear of the Acquisition Trunk Cable. This is visible as a damaged sheath or exposed wires. If a damaged Acquisition Trunk Cable is not recognized and removed from service, the cable can overheat with continued use. If a user incorrectly places the Acquisition Module into the Acquisition Module Holder, it can result in excessive bending and wear of the Acquisition Trunk Cable. This is visible as a damaged sheath or exposed wires. If a damaged Acquisition Trunk Cable is…

How do I know if I have the recalled product?

Match the codes printed on your package against the notice: UPC 00840682125499; model number 2030360-001, VU4 — UDI-DI. If the code on your package is not on that list, your unit is outside the recall.

What should I do with it right now?

Stop using it and unplug it from the wall. If it runs on a battery pack, stop charging it.

Can I get my money back?

A refund means the purchase price back, sometimes prorated by age for durable goods. Ask how they calculated it if the figure looks low.

Where was the recalled product sold?

It was distributed nationwide.