GE Medical Systems Information Technologies Inc GE HealthCare Trade IN Carescape recall: GE HealthCare has become aware of an unlikely situation where potential loss
GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
The hazard: Device safety concern
The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
What the notice says: GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-2246-2026
- Published
- June 3, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- GE Medical Systems Information Technologies Inc
- Product type
- Medical devices
- Units affected
- 2366 units
- Made in
- United States
- Lot / code info
- No UDI, Serial Number or Sales Order Number: KSA7536146, KSA7536158, KSA7536159, KSA7550156, KSA7550158, KSA7550159, KSA7667317, KSA7754389, KSA7769597, KSA7769610, KSA7769612, KSA7773420, KSA7773431, KSA7773440, KSA7120704, KSA7120767, KSA6964886, KSA6808842, KSA6808852, KSA6808866, KSA6808868, KSA6808881, KSA6808883, KSA6808912, KSA6808928, KSA6866974, KSA6664007, KSA6866963, KSA7120766, KSA7162586, KSA6623197, KSA6623216, KSA6623223, KSA6623226, KSA6623239, KSA6623248, KSA6623270, KSA6623278, KSA7391460, KSA6477094, KSA6477095, KSA6477135, KSA6477155, KSA6477159, KSA6477206, KSA6954245, KSA6954269, KSA6954276, KSA6954277, KSA6954284, KSA6954291, KSA6477096, KSA6477154, KSA7702196, KSA7702198, KSA7702199, KSA7702208, KSA7702222, KSA6623191, KSA6623194, KSA6623237, KSA6623251, KSA6623267, KSA6626626, KSA6626632, KSA6626639, KSA6626666, KSA6626670, KSA6626716, KSA6536442, KSA6536443, KSA6536454, KSA6536460, KSA6536469, KSA6536484, KSA6536485, KSA6536490, KSA6536496, KSA6536504, KSA6536526, KSA6536528, KSA6536532, KSA7120699, KSA7120728, KSA7120764, KSA7120773, KSA7162570, KSA7162593, KSA7162610, KSA7162622, KSA7162640, KSA6623198, KSA6623202, KSA6623232, KSA6623236, KSA6623275, KSA6623280, KSA6626622, KSA6626625, KSA6626635, KSA6626636, KSA6626648, KSA6626655, KSA6626659, KSA6626661, KSA6536520, KSA6536525, KSA7304431, KSA6866921, KSA6866926, KSA6866927, KSA6866983, KSA6808882, KSA6649195, KSA6649250, KSA7162575, KSA7162588, KSA7162608, KSA7162614, KSA7162656, KSA6899015, KSA6899022, KSA6808879, KSA5646278, KSA6017318, KSA6017321, KSA6017325, KSA6017333, KSA6017336, KSA6017352, KSA6017360, KSA6017361, KSA6017367, KSA6017368, KSA6017370, KSA6017371, KSA6017377, KSA6017378, KSA6017382, KSA6017383, KSA6017384, KSA6017385, KSA6017387, KSA6017388, KSA6017391, KSA6017404, KSA6017405, KSA6017406, KSA6017407, KSA6017409, KSA6017410, KSA6017416, KSA6017417, KSA6017424, KSA6017445, KSA6017450, KSA6017458, KSA6017471, KSA6017504, KSA6899026, KSA7676874, KSA7676878, KSA7676893, KSA7013925, KSA6808863, KSA6808864, KSA7304413, KSA7304437, KSA6945205, KSA6945208, KSA6945219, KSA6945223, KSA6945230, KSA6945236, KSA6945296, KSA6945314, KSA6945321, KSA6945340, KSA6945350, KSA6945353, KSA6945382, KSA6945387, KSA6899004, KSA6664048, KSA6664049, KSA6664054, KSA6664059, KSA6664075, KSA6664037, KSA6664045, KSA6664065, KSA6664069, KSA6623192, KSA6626641, KSA6626649, KSA6626687, KSA6626691, KSA6626701, KSA6626706, KSA6626707, KSA6626715, KSA6626718, KSA7120720, KSA7162574, KSA7162580, KSA7162632, KSA7162655, KSA6419987, KSA6419993, KSA6420030, KSA6420053, KSA6420080, KSA6420081, KSA6420082, KSA6420102, KSA6664002, KSA6664005, KSA6664036, KSA6664052, KSA6664057, KSA6664078, KSA6664090, KSA6664091, KSA6664006, KSA6866903, KSA6866924, KSA6866928, KSA6866933, KSA6866942, KSA6866977, KSA6866981, KSA6866987, KSA6866992, KSA6866996, KSA6899000, KSA6899060, KSA6945146, KSA6945185, KSA6945190, KSA6945191, KSA6945224, KSA6945250, KSA6945278, KSA6945294, KSA6945311, KSA6945391, KSA6945199, KSA6945204, KSA6945226, KSA6945244, KSA6945258, KSA6945302, KSA6945303, KSA6945319, KSA6945355, KSA6945358, KSA6945359, KSA6945384, KSA7013899, KSA7013902, KSA7013903, KSA7013905, KSA7013912, KSA7013914, KSA7013918, KSA7013921, KSA7013932, KSA7013940, KSA7013955, KSA7013956, KSA7013957, KSA7013963, KSA7013964, KSA7013973, KSA7013975, KSA7013986, KSA6569809, KSA6569821, KSA6569833, KSA6569840, KSA6569867, KSA6569890, KSA6569898, KSA6623189, KSA6623195, KSA6626637, KSA6626651, KSA6626690, KSA6626699, KSA6626713, KSA6626719, KSA6898999, KSA6899011, KSA6899013, KSA6899018, KSA6899021, KSA6899028, KSA6899029, KSA6899033, KSA6899034, KSA6899037, KSA6899051, KSA6899054, KSA6899059, KSA6899067, KSA6899070, KSA6899072, KSA6899074, KSA6899078, KSA6899080, KSA6899084, KSA6899085, KSA6899086, KSA6899092, KSA6899093, KSA6899097, KSA6945099, KSA6945100, KSA6945108, KSA6945113, KSA6945130, KSA6945140, KSA6945171, KSA6945175, KSA6945177, KSA6945183, KSA6945187, KSA6945189, KSA6945254, KSA694…
Products covered by this recall
GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L - NORTH AMERICA, REF #5697603-902, Product Code KSA; used in conjunction wtih ApexPro CARESCAPE Telemetry Server (CTS) v6.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Check your product
The notice names 2 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
Model or item number
On the rating label, the box or the instruction sheet.
- 5697603-902
Package size
Match the net weight or the piece count on the front of the pack.
- 2250 L
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
GE Medical Systems Information Technologies Inc — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
Who pays for a device recall
The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.
- If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
- Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
- If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
- Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.
Company contact: GE Medical Systems Information Technologies Inc — 3114 N Grandview Blvd, Waukesha, WI, 53188-1615
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was GE Medical Systems Information Technologies Inc medical… recalled?
GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station). GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: model number 5697603-902. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
Related recalls
-
Class II · last month
GE Medical Systems Information Technologies Inc Carescape Central Station CSCS V3 recall: Device may not resume patient monitoring for at least 3 minutes following
Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of…
-
Class II · last month
GE Medical Systems Information Technologies Inc Portrait Core Services Software recall: Software issue may result in a loss of patient monitoring at the Hub
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory…
-
Class II · last month
GE Medical Systems Information Technologies Inc Portrait Compute AND Core Services recall: Software issue may result in a loss of patient monitoring at the Hub
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory…
-
Class II · last month
GE Medical Systems Information Technologies Inc FRU recall: Software issue may result in a loss of patient monitoring at the Hub
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory…
-
Class II · last month
GE Medical Systems Information Technologies Inc Portrait Core Services Software V1.1 recall: Software issue may result in a loss of patient monitoring at the Hub
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory…
-
Class II · 2 months ago
GE Medical Systems Information Technologies Inc GE Healthcare MAC VU360, REF# recall: Software issue issue with the MAC VU360 v3.0 system. After selecting
Software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close…