GE Healthcare LLC SIGNA Premier systems recall: Whole-Body MR scanners used to produce images of the inside of the human body
Whole-Body MR scanners used to produce images of the inside of the human body to aid diagnosis of disease may have been manufactured/serviced with ferrous (steel) fittings instead of non-ferrous (brass) fittings on the magnet rear and if system is accessed to perform service while the magnet is ramped, the magnetic field could attract the ferrous fittings potentially resulting in personnel injury.
The hazard: High-powered magnet hazard
Loose high-powered magnets are the hazard here. The danger comes from swallowing more than one, because they then find each other through the gut wall.
What the notice says: Whole-Body MR scanners used to produce images of the inside of the human body to aid diagnosis of disease may have been manufactured/serviced with ferrous (steel) fittings instead of non-ferrous (brass) fittings on the magnet rear and if system is accessed to perform service while the magnet is ramped, the magnetic field could attract the ferrous fittings potentially resulting in personnel injury.
Why it matters: Loose high-powered magnets are dangerous when more than one is swallowed. They attract each other through the walls of the intestine and pinch the tissue trapped between them, an injury that usually needs surgery.
Symptoms and when to call a doctor
If you think a child has swallowed magnets — even if they seem entirely well — go to an emergency department and say magnets are involved. Early symptoms are indistinguishable from a stomach bug: stomach pain, vomiting, poor appetite. A single magnet may pass on its own; two or more rarely do, and the team needs to know how many are unaccounted for.
This page is general information, not medical advice. If you feel unwell, a clinician who can see you is the only person able to judge your case.
Recall details
- Recall number
- Z-1767-2026
- Published
- April 8, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- GE Healthcare LLC
- Product type
- Medical devices
- Units affected
- 41
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- UDI-DI/GTIN: 00840682135269, 00195278010797. System ID: 415353UMR2, 415723SHMR19, 413253MR3T, 908725MR, 609652MR3T, M48823, 7168453T, 513584MR5, 513636EASTMR, 513636T1MR, 605333MRPRM, 832522CRMR1, 703858PRE2, 804675VA3T, 082427110162, 082427220143, 082427020147, 082427200242, 082427160461, 082427250139, 082427140382, M4216469, 353745MR01, GON5655515, KRMR56474971, KRMR56203791, KRMR55433661, KRMR56344161, KRMR55774051, KRMR55914661, NOMRI26833321, 15974045, GON1882380, ESMRI27056983, ESMRI26960129, A203MR06, 690120MR01, 11524261, VNMRI26792525, VNMRI26592248, VNMRI26809677
- UPC codes
- 00840682135269, 082427110162, 082427220143, 082427020147, 082427200242, 082427160461, 082427250139, 082427140382
Products covered by this recall
SIGNA Premier systems
Check your product
The notice names 8 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 00840682135269
- 082427110162
- 082427220143
- 082427020147
- 082427200242
- 082427160461
- 082427250139
- 082427140382
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Magnets
GE Healthcare LLC — Loose high-powered magnets are the hazard here. The danger comes from swallowing more than one, because they then find each other through the gut wall.
- Put the product out of reach and count the magnets against the number the instructions say it should have.
- If any are unaccounted for and a child could have reached them, go to an emergency department and tell them magnets may have been swallowed.
- Make sure older children know that using magnets as fake piercings is a recognised route to a serious injury.
- Follow the remedy described on this page rather than keeping the item with the magnets glued back in.
Claiming your refund, repair or replacement
The recalling company decides the remedy and funds it. You should never pay for a recall repair or for return shipping.
- Use the company contact on this page first. Most firms run a dedicated recall page with a claim form that is far quicker than their main support line.
- If the company stops responding, escalate to the CPSC. Firms under a recall order have reporting duties to the agency.
- Have the model number, the date code and your photos ready. Many companies issue store credit by default — ask for cash or a card reversal if you would rather have it.
- No receipt? Say so. Recall refunds are routinely paid without proof of purchase, because the hazard is what triggers the obligation.
Company contact: GE Healthcare LLC — 3001 W Radio Dr, Florence, SC, 29501-7806
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was GE Healthcare LLC medical devices recalled?
Whole-Body MR scanners used to produce images of the inside of the human body to aid diagnosis of disease may have been manufactured/serviced with ferrous (steel) fittings instead of non-ferrous (brass) fittings on the magnet rear and if system is accessed to perform service while the magnet is ramped, the magnetic field could attract the ferrous fittings potentially resulting in personnel injury. Whole-Body MR scanners used to produce images of the inside of the human body to aid diagnosis of disease may have been manufactured/serviced with ferrous (steel) fittings instead of non-ferrous…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 00840682135269, 082427110162, 082427220143. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Put the product out of reach and count the magnets against the number the instructions say it should have.
Can I get my money back?
Use the company contact on this page first. Most firms run a dedicated recall page with a claim form that is far quicker than their main support line.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · last month
GE Healthcare Centricity PACS Server V7, intended for managing digital medical images recall: Under certain conditions historical imaging data migrated from a legacy
Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this…
-
Class II · last month
GE Healthcare Centricity PACS V2, V3, V4, V6, V7, software product is intended recall: Under certain conditions historical imaging data migrated from a legacy
Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this…
-
Class II · last month
GE Healthcare Centricity Enterprise Archive V2, V3, and V4 recall: Under certain conditions historical imaging data migrated from a legacy
Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this…
-
Class II · last month
GE Healthcare Core Services Hub V8, a set of backend services including Cross recall: Under certain conditions historical imaging data migrated from a legacy
Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this…
-
Class II · last month
GE Healthcare Workflow Core Services V8, intended to provide set of services recall: Under certain conditions historical imaging data migrated from a legacy
Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this…
-
Class II · last month
GE Healthcare Enterprise Archive V8, software product for receiving, archiving recall: Under certain conditions historical imaging data migrated from a legacy
Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this…