Draeger, Inc. Brand Name: Atlan Product Name: Atlan A350 Model/ recall: The device either indicated a failure of the piston ventilator before use
The device either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
The hazard: Device safety concern
The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
What the notice says: The device either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-2616-2025
- Published
- October 8, 2025
- Risk level
- Class I
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Draeger, Inc.
- Product type
- Medical devices
- Units affected
- 530
- Distribution
- Alabama, Arizona, California, Connecticut, Florida, Georgia, Idaho, Illinois, Indiana, Kansas, Kentucky, Massachusetts, Minnesota, Missouri, North Carolina, Nebraska, New Mexico, Nevada, New York, Ohio, Pennsylvania, Texas, Utah, Virginia, Wisconsin
- Made in
- United States
- Lot / code info
- Lot Code: UDI-DI Number: 04048675556176 Serial Numbers Distributed in the US: ASTB-0181 ASTC-0005,ASTC-0006,ASTC-0007 ASTD-0244,ASTD-0259 ASTE-0180,ASTE-0254,ASTE-0255 ASTH-0083,ASTH-0464 ASUA-0479,ASUA-0482 ASUB-0012,ASUB-0013,ASUB-0015,ASUB-0017,ASUB-0018,ASUB-0292 ASUC-0065,ASUC-0084,ASUC-0085,ASUC-0086,ASUC-0087,ASUC-0088,ASUC-0090,ASUC-0091, ASUC-0092,ASUC-0096,ASUC-0097,ASUC-0099,ASUC-0100,ASUC-0101,ASUC-0102,ASUC-0103, ASUC-0104,ASUC-0105,ASUC-0107,ASUC-0108,ASUC-0109,ASUC-0110,ASUC-0111,ASUC-0112, ASUC-0113,ASUC-0116,ASUC-0117,ASUC-0118,ASUC-0119,ASUC-0120,ASUC-0121,ASUC-0122, ASUC-0123,ASUC-0125,ASUC-0126,ASUC-0127,ASUC-0128,ASUC-0129,ASUC-0130,ASUC-0131, ASUC-0132,ASUC-0133,ASUC-0134,ASUC-0135,ASUC-0136,ASUC-0137,ASUC-0138,ASUC-0139, ASUC-0140,ASUC-0141,ASUC-0142,ASUC-0143,ASUC-0143,ASUC-0144,ASUC-0145,ASUC-0162, ASUC-0163,ASUC-0164,ASUC-0165,ASUC-0166,ASUC-0167,ASUC-0168,ASUC-0169,ASUC-0170, ASUC-0171,ASUC-0172,ASUC-0173,ASUC-0203,ASUC-0204,ASUC-0205,ASUC-0206,ASUC-0207, ASUC-0208,ASUC-0209,ASUC-0210,ASUC-0218,ASUC-0219,ASUC-0220,ASUC-0221,ASUC-0222, ASUC-0223,ASUC-0224,ASUC-0225,ASUC-0235,ASUC-0236,ASUC-0237,ASUC-0340,ASUC-0341 ASUD-0244,ASUD-0261,ASUD-0262,ASUD-0263,ASUD-0267,ASUD-0281,ASUD-0282,ASUD-0283 Serial Numbers Distributed OUS ASUA-0353,ASUA-0441,ASUA-0447,ASUA-0472,ASUA-0478,ASUA-0496,,SUA-0497,ASUA-0500, ASUB-0004,ASUB-0024,ASUB-0025,ASUB-0030,ASUB-0033,ASUB-0034,ASUB-0035,ASUB-0038, ASUB-0039,ASUB-0040,ASUB-0041,ASUB-0043,ASUB-0044,ASUB-0046,ASUB-0048,ASUB-0063, ASUB-0064,ASUB-0066,ASUB-0067,ASUB-0069,ASUB-0073,ASUB-0074,ASUB-0075,ASUB-0076, ASUB-0077,ASUB-0095,ASUB-0096,ASUB-0097,ASUB-0098,ASUB-0099,ASUB-0100,ASUB-0101, ASUB-0102,ASUB-0104,ASUB-0105,ASUB-0107,ASUB-0108,ASUB-0109,ASUB-0112,ASUB-0113, ASUB-0114,ASUB-0115,ASUB-0116,ASUB-0118,ASUB-0120,ASUB-0121,ASUB-0122,ASUB-0123, ASUB-0124,ASUB-0128,ASUB-0129,ASUB-0130,ASUB-0136,ASUB-0137,ASUB-0139,ASUB-0140, ASUB-0141,ASUB-0143,ASUB-0144,ASUB-0145,ASUB-0146,ASUB-0147,ASUB-0148,ASUB-0149, ASUB-0150,ASUB-0151,ASUB-0152,ASUB-0153,ASUB-0154,ASUB-0155,ASUB-0156,ASUB-0157, ASUB-0158,ASUB-0159,ASUB-0160,ASUB-0161,ASUB-0162,ASUB-0163,ASUB-0177,ASUB-0178, ASUB-0188,ASUB-0189,ASUB-0190,ASUB-0191,ASUB-0192,ASUB-0193,ASUB-0194,ASUB-0195, ASUB-0196,ASUB-0197,ASUB-0198,ASUB-0199,ASUB-0201,ASUB-0202,ASUB-0203,ASUB-0204, ASUB-0205,ASUB-0206,ASUB-0207,ASUB-0208,ASUB-0209,ASUB-0210,ASUB-0211,ASUB-0223, ASUB-0224,ASUB-0225,ASUB-0226,ASUB-0227,ASUB-0229,ASUB-0230,ASUB-0231,ASUB-0232, ASUB-0233,ASUB-0234,ASUB-0235,ASUB-0236,ASUB-0237,ASUB-0238,ASUB-0240,ASUB-0243, ASUB-0245,ASUB-0252,ASUB-0253,ASUB-0254,ASUB-0255,ASUB-0256,ASUB-0257,ASUB-0258, ASUB-0260,ASUB-0261,ASUB-0262,ASUB-0263,ASUB-0264,ASUB-0265,ASUB-0271,ASUB-0272, ASUB-0273,ASUB-0276,ASUB-0278,ASUC-0001ASUC-0004,ASUC-0016,ASUC-0017,ASUC-0018, ASUC-0019,ASUC-0020,ASUC-0022,ASUC-0024,ASUC-0026,,ASUC-0028,,ASUC-0030,ASUC-0031, ASUC-0032,ASUC-0033,ASUC-0034,ASUC-0035,ASUC-0036,ASUC-0039,ASUC-0061,ASUC-0062, ASUC-0063,ASUC-0226,ASUC-0227,ASUC-0228,ASUC-0229,ASUC-0230,ASUC-0232,ASUC-0233, ASUC-0234,ASUC-0247,ASUC-0248,ASUC-0249,ASUC-0250,ASUC-0251,ASUC-0252,ASUC-0253, ASUC-0254,ASUC-0255,ASUC-0256,ASUC-0257,ASUC-0262,ASUC-0263,ASUC-0264,ASUC-0265, ASUC-0266,ASUC-0267,ASUC-0269,ASUC-0272,ASUC-0273,ASUC-0274,ASUC-0275,ASUC-0276, ASUC-0277,ASUC-0278,ASUC-0279,ASUC-0284,ASUC-0286,ASUC-0287,ASUC-0288,ASUC-0289, ASUC-0290,ASUC-0291,ASUC-0292,ASUC-0293,ASUC-0294,ASUC-0295,ASUC-0296,ASUC-0300, ASUC-0301,ASUC-0302,ASUC-0303,ASUC-0304,ASUC-0305,ASUC-0306ASUC-0307,ASUC-0308, ASUC-0309,ASUC-0310,ASUC-0311,ASUC-0314,ASUC-0315,ASUC-0316,ASUC-0319,ASUC-0320, ASUC-0321,ASUC-0322,ASUC-0323,ASUC-0339,ASUC-0342,ASUC-0343,ASUC-0344,ASUC-0345, ASUC-0346,ASUC-0348,ASUC-0349,ASUC-0350,ASUC-0351,ASUC-0352,ASUC-0353,ASUC-0354, ASUC-0355,ASUC-0356,ASUC-0357,ASUC-0358,ASUC-0359,ASUC-0360,ASUC-0361,ASUC-0362, ASUC-0363,ASUC-0364,ASUC-0365,ASUC-0366,ASUC-0367,ASUC-0380,ASUC-0381,ASUC-0382, ASUC-0383,ASUC-0384,ASUC-0385,ASUC-0386,ASUC-0387,ASUC-03…
- UPC codes
- 04048675556176
Products covered by this recall
Brand Name: Atlan Product Name: Atlan A350 Model/Catalog Number: 8211500 Software Version: Not applicable Product Description: The Atlan is an anesthesia workstation intended for use in anesthetizing adults, pediatrics, and neonates. The device can be used for mechanical ventilation, manual ventilation, pressure-supported spontaneous breathing, and spontaneous breathing. The device is specified for inhalational anesthesia and/or patient ventilation in accordance with the intended use during surgical or diagnostic interventions Component: Not applicable
Check your product
The notice names 5 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 04048675556176
Model or item number
On the rating label, the box or the instruction sheet.
- 8211500 Software Version
Serial number
On the rating plate, often next to the date of manufacture.
- Distributed OUS ASUA-0353
- ASUA-0441
- ASUA
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Draeger, Inc. — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
Who pays for a device recall
The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.
- If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
- Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
- If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
- Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.
Company contact: Draeger, Inc. — 3135 Quarry Rd, Telford, PA, 18969-1042
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Draeger, Inc. medical devices recalled?
The device either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use. The device either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 04048675556176; model number 8211500 Software Version. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
Which states were affected?
The notice lists Alabama, Arizona, California, Connecticut, Florida, Georgia, Idaho, Illinois, Indiana, Kansas, Kentucky, Massachusetts and others.
Related recalls
-
Class I · 18 days ago
Draeger, Inc. VentStar Resus Neo. Neonatal. 25 Units per Box recall: Potential cracks in the hose, which could lead to leakage or the hose falling
Potential cracks in the hose, which could lead to leakage or the hose falling off.
-
Class I · 5 months ago
Draeger, Inc. Atlan A350XL recall: The Atlan anesthesia workstation either indicated a failure of the piston
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation…
-
Class I · 5 months ago
Draeger, Inc. Atlan A350 recall: The Atlan anesthesia workstation either indicated a failure of the piston
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation…
-
Class I · 10 months ago
Draeger Vapor 2000. Unheated, calibrated anesthetic vaporizer recall: A certain component of affected devices was not delivered within
A certain component of affected devices was not delivered within specification and contained impurities.
-
Class I · 10 months ago
Draeger Vapor 3000. Unheated, calibrated anesthetic vaporizer recall: A certain component of affected devices was not delivered within
A certain component of affected devices was not delivered within specification and contained impurities.
-
Class I · 10 months ago
Draeger, Inc. ErgoStar CM 60, Model/ recall: Multiple complaints were reported in which cracks formed in the hose
Multiple complaints were reported in which cracks formed in the hose of specific catheter mounts.