Datex-Ohmeda, Inc. Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system recall: Anesthesia Delivery Systems containing certain electronic gas mixer
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.
The hazard: Device safety concern
The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
What the notice says: Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-3191-2026
- Published
- September 23, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Datex-Ohmeda, Inc.
- Product type
- Medical devices
- Units affected
- 7 units
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- Serial Numbers: ANAS01357 ANAP00609 ANAQ00988 ANAS00586 ANAR00543 ANAP01088 ANAM00443
Products covered by this recall
Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000
What to do about Device
Datex-Ohmeda, Inc. — The manufacturer or the FDA identified a problem with this device. The model, lot and catalogue numbers are what define the scope.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
- Ask the supplier or clinic that provided it what the remedy is in your case.
- If you are a facility, quarantine affected stock and log the correction.
Who pays for a device recall
The manufacturer funds the remedy. The practical question is usually which link in the chain executes it.
- If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
- Ask for written confirmation that the remedy was applied to your serial number. That is what an insurer or a future clinician will want to see.
- If you bought it yourself online or from a pharmacy, go to the manufacturer’s recall page rather than the retailer; device claims are handled centrally.
- Where the remedy is a software update, there is nothing to claim, but do confirm the version changed afterwards.
Company contact: Datex-Ohmeda, Inc. — 3030 Ohmeda Dr, Madison, WI, 53718-6704
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Datex-Ohmeda, Inc. medical devices recalled?
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect. Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically…
How do I know if I have the recalled product?
The notice does not publish package codes for this recall. Compare the product name, size and the purchase window against the official FDA — Device enforcement record, and contact the company if you are unsure.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
If a clinic fitted or supplied the device, start there — they hold the serial-number records the manufacturer needs.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · last month
Datex-Ohmeda, Inc. Et Control, Software option for anesthesia gas machine to achieve recall: Anesthesia Delivery Systems containing certain electronic gas mixer
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2…
-
Class II · last month
Datex-Ohmeda, Inc. Aisys CS2, Datex-Ohmeda anesthesia system recall: Anesthesia Delivery Systems containing certain electronic gas mixer
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2…
-
Class II · last month
Datex-Ohmeda, Inc. Aisys, Datex-Ohmeda anesthesia system recall: Anesthesia Delivery Systems containing certain electronic gas mixer
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2…
-
Class I · 4 months ago
Datex-Ohmeda Inc Giraffe Stand-alone Infant Resuscitation System recall: Blender knob shaft can loosen, which can affect the delivered oxygen
Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not…
-
Class I · 4 months ago
Datex-Ohmeda Inc Giraffe Warmer with Integrated Resuscitation System recall: Blender knob shaft can loosen, which can affect the delivered oxygen
Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not…
-
Class I · 4 months ago
Datex-Ohmeda Inc Panda Freestanding Warmer, Infant Radiant Warmer Product Identifiers recall: Blender knob shaft can loosen, which can affect the delivered oxygen
Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not…