CareFusion 303, Inc. BD Alaris recall: If infusion pump is dropped or severely jarred this may damage the pump
If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested, and inspected by qualified service personnel prior to reuse; tip sheet, cleaning and disinfecting procedure and reference guide updated.
The hazard: Device safety concern
This is a device enforcement record. Check whether the remedy is a correction, a software update or a removal before changing anything.
What the notice says: If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested, and inspected by qualified service personnel prior to reuse; tip sheet, cleaning and disinfecting procedure and reference guide updated.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-0430-2026
- Published
- November 26, 2025
- Risk level
- Class I
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- CareFusion 303, Inc.
- Product type
- Medical devices
- Units affected
- 2,574,396
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- All Alaris Pump Modules distributed after June 19, 2004. REF/UDI-DI: 8100/10885403810015, 10885403810046, 10885403810039, 10885403222054, 10885403517723, 10885403213212, 10885403211423; 49000437/10885403471018; 49000270/10885403247743
- UPC codes
- 10885403810015, 10885403810046, 10885403810039, 10885403222054, 10885403517723, 10885403213212, 10885403211423, 10885403471018, 10885403247743
Products covered by this recall
BD Alaris infusion Pump Module 8100, REF: 8100; KIT, BEZEL ASSY 8100BD, REF: 49000437; KIT ASSY ONE PIECE BEZEL 8100, REF: 49000270. Used with the following new and updated documents: BD Alaris Infusion System Inspection Safety Reminder, Part: BD-163124, Revision: 00; Cleaning and Disinfecting Procedure, Part: P00000864; Cleaning and Disinfecting Quick Reference Guide
Check your product
The notice names 13 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 10885403810015
- 10885403810046
- 10885403810039
- 10885403222054
- 10885403517723
- 10885403213212
- 10885403211423
- 10885403471018
- 10885403247743
Model or item number
On the rating label, the box or the instruction sheet.
- 8100
- 49000437
- BD-163124
- P00000864
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
CareFusion 303, Inc. — This is a device enforcement record. Check whether the remedy is a correction, a software update or a removal before changing anything.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
Replacements, repairs and reimbursement
Device remedies run through the supply chain, so the route depends on who sold you the unit.
- Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
- If the device was billed to insurance or Medicare, the replacement is usually billed the same way; you should not face a second deductible for a recall.
- Keep every piece of correspondence. If a recalled device caused harm, that paper trail is what a claim rests on.
- Consumer-purchased devices such as monitors and test kits are normally replaced free, with return shipping paid.
Company contact: CareFusion 303, Inc. — 10020 Pacific Mesa Blvd, San Diego, CA, 92121-4386
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was CareFusion 303, Inc. medical devices recalled?
If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested, and inspected by qualified service personnel prior to reuse; tip sheet, cleaning and disinfecting procedure and reference guide updated. If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 10885403810015, 10885403810046, 10885403810039; model number 8100, 49000437, BD-163124. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 5 months ago
CareFusion 303, Inc. BD Pyxis MedBank Tower REF: 1145-00 Medication cabinet recall: Fingerprint scanner may overheat to a temperature to cause 1st degree burn
Fingerprint scanner may overheat to a temperature to cause 1st degree burn.
-
Class II · 5 months ago
CareFusion 303, Inc. BD Pyxis Mini Main REF: 349 Medication cabinet recall: Fingerprint scanner may overheat to a temperature to cause 1st degree burn
Fingerprint scanner may overheat to a temperature to cause 1st degree burn.
-
Class II · 5 months ago
CareFusion 303, Inc. BD Pyxis Anesthesia Station 4000 REF: 338 Medication cabinet recall: Fingerprint scanner may overheat to a temperature to cause 1st degree burn
Fingerprint scanner may overheat to a temperature to cause 1st degree burn.
-
Class II · 5 months ago
CareFusion 303, Inc. BD Pyxis MedStation ES REF: 323 Medication cabinet recall: Fingerprint scanner may overheat to a temperature to cause 1st degree burn
Fingerprint scanner may overheat to a temperature to cause 1st degree burn.
-
Class II · 5 months ago
CareFusion 303, Inc. BD Pyxis Anesthesia Station ES REF: 327 Medication cabinet recall: Fingerprint scanner may overheat to a temperature to cause 1st degree burn
Fingerprint scanner may overheat to a temperature to cause 1st degree burn.
-
Class II · 5 months ago
CareFusion 303, Inc. BD Pyxis MedStation 4000 REF: 303 Medication cabinet recall: Fingerprint scanner may overheat to a temperature to cause 1st degree burn
Fingerprint scanner may overheat to a temperature to cause 1st degree burn.