Brand Name: Olympus ShockPulse-SE Lithotripsy System with Generator recall: Mis-wired component-the improperly wired component can result in additional
Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards and may decrease the overall reliability of the system and result in delay of treatment. It was discovered that additional serial numbers of ShockPulse generators may also have the mis-wired component. As a result, this is expanding the action to the additional impacted devices.
The hazard: Device safety concern
A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.
What the notice says: Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards and may decrease the overall reliability of the system and result in delay of treatment. It was discovered that additional serial numbers of ShockPulse generators may also have the mis-wired component. As a result, this is expanding the action to the additional impacted devices.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-1494-2026
- Published
- March 11, 2026
- Risk level
- Class II
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Olympus Corporation of the Americas
- Product type
- Medical devices
- Units affected
- 55 units (All OUS)
- Made in
- United States
- Lot / code info
- Model/Catalog Number: SPL-SR, contains generator SPL-G; UDI: SPL-SR 00821925043824, SPL-G 00821925044203; Serial Numbers: CG5026, CG5027, CG5028, CG5029, CG5030, CG5031, CG5032, CG5033, CG5034, CG5035, CG5036, CG5037, CG5039, CG5040, CG5041, CG5042, CG5043, CG5045, CG5046, CG5047, CG5048, CG5049, CG5050, CG5051, CG5052, CG5053, CG5054, CG5055, CG5056, CG5057, CG5058, CG5059, CG5060, CG5061, CG5062, CG5063, CG5065, CG5066, CG5067, CG6037, CG6061, CG6069, CG6070, CG6071, CG6088, CG6094, CG6095, CG6096, CG6097, CG6098, CG6099, CG7000, CG7001, CG7002, CG7003;
- UPC codes
- 00821925043824, 00821925044203
Products covered by this recall
Brand Name: Olympus ShockPulse-SE Lithotripsy System with Generator Product Name: ShockPulse-SE Lithotripsy System - Reuseable Probes Model/Catalog Number: SPL-SR, contains generator SPL-G Product Description: An assembly of devices that uses a combination of, or individually applied, pneumatically-created ballistic shock waves and ultrasonic waves for the intracorporeal fragmentation of stones (calculi) in the urinary tract (i.e., kidney, ureter, and bladder) for their removal. It typically includes a pressure and power regulator/control unit, handpieces, and two types of probes of various sizes. The probes are attached to the two handpieces which are connected to the control unit that supplies the energy for both the pneumatic and ultrasonic functions. Energy is applied to a calculus which is fragmented in situ and removed via a suction system facilitated by the lithotripsy system.
Check your product
The notice names 9 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 00821925043824
- 00821925044203
Model or item number
On the rating label, the box or the instruction sheet.
- SPL-SR
- contains generator SPL-G
Serial number
On the rating plate, often next to the date of manufacture.
- CG5026
- CG5027
- CG5028
- CG5029
- CG5030
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Olympus Corporation of the Americas — A device recall was filed for this product. For anything implanted or relied on daily, the clinician comes before the paperwork.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
Claiming on a recalled device
Device recalls are tracked by serial and lot number, so your claim moves faster if you lead with those.
- Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.
- Ask explicitly whether your unit is in the affected range; many recalls cover only part of a model line.
- If you need the device daily, ask what interim arrangement they offer while the remedy is carried out.
- Log the date, the name and the reference number of every call. Device recalls can run for months.
Company contact: Olympus Corporation of the Americas — 3500 Corporate Pkwy, Center Valley, PA, 18034-8229
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Olympus Corporation of the Americas medical devices recalled?
Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards and may decrease the overall reliability of the system and result in delay of treatment. It was discovered that additional serial numbers of ShockPulse generators may also have the mis-wired component. As a result, this is expanding the action to the additional impacted devices. Mis-wired component-the improperly wired component can result in additional noise on the power supply…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 00821925043824, 00821925044203; model number SPL-SR, contains generator SPL-G. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
Have the model, catalogue, lot and serial numbers ready before you call — they are on the device label or its box.
Related recalls
-
Class II · 18 days ago
Olympus M3 Gold Autoclavable Foroblique Telescope 30 recall: There are identified discrepancies between the sterilization parameters
There are identified discrepancies between the sterilization parameters listed in the Instructions for Use and the sterilization parameters listed in…
-
Class II · 18 days ago
Olympus Tripp-Lite P580-025 25ft DisplayPort Cable recall: While in the high-power operating mode, a video dropout/blackout may occur
While in the high-power operating mode, a video dropout/blackout may occur when using a particular accessory cable.
-
Class II · 18 days ago
Olympus M3 Gold Autoclavable Lateral Telescope 70 recall: There are identified discrepancies between the sterilization parameters
There are identified discrepancies between the sterilization parameters listed in the Instructions for Use and the sterilization parameters listed in…
-
Class II · 25 days ago
Olympus BiCOAG Hemostasis Probe 7Fr Fixed Pin Connector. Model/ recall: Potential for detachment of ceramic tip due to manufacturing without
Potential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive.
-
Class II · 25 days ago
Olympus BiCOAG Hemostasis Probe 10Fr Coaxial Connector. Model/ recall: Potential for detachment of ceramic tip due to manufacturing without
Potential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive.
-
Class II · 25 days ago
Olympus EZDilate Endoscopic Balloon Dilator Wired-guided 11, 12, 13mm recall: Potential for the inflation chart label affixed to the shaft of the device
Potential for the inflation chart label affixed to the shaft of the device to be incorrect and indicate a balloon diameter of 18, 19, or 20 mm,