Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product recall: The affected devices are the Dexcom G7 Continuous Glucose Monitoring System

Recall published 1+ years ago

The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App software (versions 1.4.0 or earlier), henceforth referred to as the App. The affected App does not provide an expected "Sensor Failed" alert when the transmitter sends a "transmitterFailed" error message to the App when the CGM experiences a hardware or firmware failure. Instead, the App ends the CGM sensor session, stops reporting glucose values and displays the Start Sensor screen or No active sensor message without alerting the user, which can lead to missed detection of a hyperglycemic or hypoglycemic event and a delay in treatment.

The hazard: Device safety concern

This is a device enforcement record. Check whether the remedy is a correction, a software update or a removal before changing anything.

What the notice says: The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App software (versions 1.4.0 or earlier), henceforth referred to as the App. The affected App does not provide an expected "Sensor Failed" alert when the transmitter sends a "transmitterFailed" error message to the App when the CGM experiences a hardware or firmware failure. Instead, the App ends the CGM sensor session, stops reporting glucose values and displays the Start Sensor screen or No active sensor message without alerting the user, which can lead to missed detection of a hyperglycemic or hypoglycemic event and a delay in treatment.

Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.

Recall details

Recall number
Z-2448-2025
Published
September 10, 2025
Risk level
Class I
Status
Ongoing
Issued by
U.S. Food and Drug Administration
Brand
Dexcom, Inc.
Product type
Medical devices
Units affected
907842
Distribution
Nationwide
Made in
United States
Lot / code info
Lot Code: UDI - SW1355- 00386270005168
UPC codes
00386270005168

Products covered by this recall

Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Number: SW13355 Software Version: versions 2.8.0 and earlier Product Description: Dexcom G7 CGM watchOS App - The watchOS app is installed along with the iOS phone app and the versions used of each are the same for each user. The watchOS app cannot be installed independently from the iOS phone app Component: Dexcom G7 CGM iOS App

Check your product

The notice names 2 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.

UPC / barcode

Read the twelve digits under the barcode on the package.

  • 00386270005168

Model or item number

On the rating label, the box or the instruction sheet.

  • SW13355 Software Version

Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.

What to do about Device

Dexcom, Inc. — This is a device enforcement record. Check whether the remedy is a correction, a software update or a removal before changing anything.

  1. Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
  2. If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
  3. Ask the supplier or clinic that provided it what the remedy is in your case.
  4. If you are a facility, quarantine affected stock and log the correction.

Replacements, repairs and reimbursement

Device remedies run through the supply chain, so the route depends on who sold you the unit.

  • Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
  • If the device was billed to insurance or Medicare, the replacement is usually billed the same way; you should not face a second deductible for a recall.
  • Keep every piece of correspondence. If a recalled device caused harm, that paper trail is what a claim rests on.
  • Consumer-purchased devices such as monitors and test kits are normally replaced free, with return shipping paid.

Company contact: Dexcom, Inc. — 6340 Sequence Dr, San Diego, CA, 92121-4356

The official notice

This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.

Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls

Questions people ask about this recall

Why was Dexcom, Inc. medical devices recalled?

The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App software (versions 1.4.0 or earlier), henceforth referred to as the App. The affected App does not provide an expected "Sensor Failed" alert when the transmitter sends a "transmitterFailed" error message to the App when the CGM experiences a hardware or firmware failure. Instead, the App ends the CGM sensor session, stops reporting glucose values and displays the Start Sensor…

How do I know if I have the recalled product?

Match the codes printed on your package against the notice: UPC 00386270005168; model number SW13355 Software Version. If the code on your package is not on that list, your unit is outside the recall.

What should I do with it right now?

Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.

Can I get my money back?

Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.

Where was the recalled product sold?

It was distributed nationwide.

Related recalls