Abiomed, Inc. Automated Impella Controller with the below product descriptions recall: Retrospective submission for following issues identified: 1. Alarm failures
Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console.
The hazard: Device safety concern
This is a device enforcement record. Check whether the remedy is a correction, a software update or a removal before changing anything.
What the notice says: Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console.
Why it matters: The FDA recorded an enforcement action against this device. Most device recalls are corrections rather than removals, so the device often stays in use while the manufacturer issues a fix.
Recall details
- Recall number
- Z-2238-2026
- Published
- June 3, 2026
- Risk level
- Class I
- Status
- Ongoing
- Issued by
- U.S. Food and Drug Administration
- Brand
- Abiomed, Inc.
- Product type
- Medical devices
- Units affected
- 625 units
- Distribution
- Nationwide
- Made in
- United States
- Lot / code info
- 1. Product Code: 0042-0000-CA. GTIN: 00813502011272. Serial Numbers: IC1324, IC1346, IC1480, IC1481, IC1543, IC1544, IC1572, IC1595, IC1598, IC1664, IC1665, IC1715. 2. Product Code: 0042-0000-EU. GTIN: 00813502011289. Serial Numbers: IC1217, IC1265, IC1275, IC1276, IC1288, IC1284, IC1301, IC1302, IC1304, IC1305, IC1344, IC1341, IC1343, IC1340, IC1345, IC1350, IC1347, IC1349, IC1383, IC1381, IC1384, IC1386, IC1393, IC1396, IC1397, IC1395, IC1413, IC1415, IC1411, IC1416, IC1490, IC1488, IC1522, IC1523, IC1524, IC1525, IC1547, IC1549, IC1546, IC1548, IC1578, IC1577, IC1579, IC1580, IC1594, IC1584, IC1590, IC1592, IC1591, IC1585, IC1588, IC1653, IC1676, IC1685, IC1684, IC1677, IC1678, IC1699, IC1700. 3. Product Code: 0042-0000-UK. GTIN: 00813502011296. Serial Numbers: IC1387, IC1389, IC1388, IC1519, IC1542, IC1545, IC1581, IC1582, IC1583, IC1573. 4. Product Code: 0042-0000-US. GTIN: 00813502010022. Serial Numbers: IC1004, IC1002, IC1001, IC1006, IC1007, IC1008, IC1014, IC1015, IC1017, IC1019, IC1024, IC1025, IC1028, IC1029, IC1026, IC1027, IC1033, IC1035, IC1031, IC1032, IC1036, IC1037, IC1038, IC1039, IC1041, IC1042, IC1044, IC1049, IC1043, IC1046, IC1053, IC1055, IC1052, IC1040, IC1030, IC1056, IC1047, IC1051, IC1050, IC1060, IC1061, IC1062, IC1064, IC1065, IC1066, IC1067, IC1074, IC1070, IC1071, IC1072, IC1073, IC1068, IC1069, IC1058, IC1075, IC1076, IC1077, IC1078, IC1079, IC1084, IC1085, IC1083, IC1080, IC1088, IC1086, IC1087, IC1081, IC1082, IC1100, IC1099, IC1093, IC1095, IC1089, IC1094, IC1096, IC1097, IC1106, IC1102, IC1105, IC1104, IC1108, IC1103, IC1107, IC1110, IC1115, IC1120, IC1111, IC1112, IC1114, IC1116, IC1117, IC1119, IC1133, IC1113, IC1132, IC1124, IC1134, IC1135, IC1136, IC1137, IC1138, IC1140, IC1141, IC1142, IC1139, IC1144, IC1143, IC1121, IC1122, IC1123, IC1147, IC1148, IC1145, IC1146, IC1149, IC1150, IC1151, IC1152, IC1153, IC1154, IC1157, IC1158, IC1159, IC1156, IC1162, IC1163, IC1164, IC1165, IC1166, IC1167, IC1168, IC1169, IC1170, IC1171, IC1174, IC1175, IC1172, IC1173, IC1178, IC1160, IC1176, IC1177, IC1179, IC1180, IC1181, IC1182, IC1186, IC1155, IC1183, IC1185, IC1184, IC1187, IC1191, IC1188, IC1189, IC1161, IC1190, IC1192, IC1193, IC1196, IC1219, IC1218, IC1194, IC1197, IC1198, IC1199, IC1213, IC1222, IC1226, IC1221, IC1225, IC1214, IC1220, IC1223, IC1224, IC1227, IC1211, IC1212, IC1208, IC1210, IC1215, IC1209, IC1202, IC1201, IC1200, IC1207, IC1203, IC1205, IC1206, IC1204, IC1235, IC1234, IC1228, IC1230, IC1231, IC1233, IC1229, IC1236, IC1237, IC1238, IC1239, IC1240, IC1242, IC1244, IC1245, IC1243, IC1250, IC1247, IC1258, IC1246, IC1251, IC1252, IC1254, IC1256, IC1257, IC1253, IC1255, IC1261, IC1264, IC1259, IC1262, IC1263, IC1269, IC1274, IC1272, IC1270, IC1271, IC1273, IC1291, IC1300, IC1296, IC1297, IC1289, IC1290, IC1293, IC1298, IC1299, IC1295, IC1292, IC1294, IC1306, IC1310, IC1309, IC1307, IC1311, IC1313, IC1314, IC1312, IC1315, IC1316, IC1317, IC1319, IC1321, IC1320, IC1322, IC1325, IC1331, IC1328, IC1330, IC1332, IC1326, IC1327, IC1333, IC1336, IC1329, IC1335, IC1342, IC1338, IC1339, IC1337, IC1334, IC1358, IC1356, IC1357, IC1353, IC1354, IC1355, IC1351, IC1352, IC1364, IC1365, IC1360, IC1361, IC1362, IC1363, IC1359, IC1379, IC1377, IC1380, IC1375, IC1376, IC1372, IC1374, IC1378, IC1373, IC1382, IC1390, IC1400, IC1408, IC1406, IC1409, IC1394, IC1403, IC1401, IC1405, IC1407, IC1402, IC1404, IC1410, IC1412, IC1414, IC1417, IC1418, IC1421, IC1422, IC1423, IC1419, IC1420, IC1425, IC1426, IC1431, IC1424, IC1429, IC1427, IC1430, IC1428, IC1440, IC1441, IC1434, IC1435, IC1436, IC1437, IC1442, IC1443, IC1432, IC1433, IC1438, IC1439, IC1450, IC1446, IC1448, IC1368, IC1444, IC1445, IC1449, IC1462, IC1461, IC1469, IC1466, IC1467, IC1470, IC1473, IC1474, IC1475, IC1465, IC1453, IC1454, IC1460, IC1471, IC1472, IC1459, IC1463, IC1464, IC1455, IC1458, IC1451, IC1452, IC1456, IC1457, IC1478, IC1484, IC1485, IC1476, IC1477, IC1479, IC1483, IC1482, IC1489, IC1491, IC1492, IC1486, IC1496, IC1495, IC1507, IC1…
- UPC codes
- 00813502011272, 00813502011289, 00813502011296, 00813502010022
Products covered by this recall
Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US.
Check your product
The notice names 30 codes you can read straight off the package. Type the one printed on yours and the list below narrows to it.
UPC / barcode
Read the twelve digits under the barcode on the package.
- 00813502011272
- 00813502011289
- 00813502011296
- 00813502010022
Model or item number
On the rating label, the box or the instruction sheet.
- 0042-0000-CA. 2. Impella Controller
- EU
- 0042-0000-EU. 3. Impella Controller
- UK
- 0042-0000-UK. 4. Impella Controller
- US
Serial number
On the rating plate, often next to the date of manufacture.
- IC1324
- IC1346
- IC1480
- IC1481
- IC1543
- IC1217
- IC1265
- IC1275
- IC1276
- IC1288
- IC1387
- IC1389
- IC1388
- IC1519
- IC1542
- IC1004
- IC1002
- IC1001
- IC1006
- IC1007
Codes are lifted verbatim from the FDA — Device enforcement notice. A code that is missing here may still be covered — the official notice is the complete list.
What to do about Device
Abiomed, Inc. — This is a device enforcement record. Check whether the remedy is a correction, a software update or a removal before changing anything.
- Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
- If you are a facility, quarantine affected stock and log the correction.
- Report device problems to the FDA through MedWatch.
- If the device is implanted or you depend on it daily, speak to your clinician before changing how you use it.
Replacements, repairs and reimbursement
Device remedies run through the supply chain, so the route depends on who sold you the unit.
- Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
- If the device was billed to insurance or Medicare, the replacement is usually billed the same way; you should not face a second deductible for a recall.
- Keep every piece of correspondence. If a recalled device caused harm, that paper trail is what a claim rests on.
- Consumer-purchased devices such as monitors and test kits are normally replaced free, with return shipping paid.
Company contact: Abiomed, Inc. — 22 Cherry Hill Dr, Danvers, MA, 01923-2575
The official notice
This page summarises a notice published by U.S. Food and Drug Administration. The agency notice is the authoritative version — read it before you act, and treat it as the final word if anything here looks different.
Open the FDA — Device enforcement record U.S. Food and Drug Administration recalls
Questions people ask about this recall
Why was Abiomed, Inc. medical devices recalled?
Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console. Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current…
How do I know if I have the recalled product?
Match the codes printed on your package against the notice: UPC 00813502011272, 00813502011289, 00813502011296; model number 0042-0000-CA. 2. Impella Controller, EU, 0042-0000-EU. 3. Impella Controller. If the code on your package is not on that list, your unit is outside the recall.
What should I do with it right now?
Match the model, catalogue, lot and serial numbers on your device against the identifiers on this page.
Can I get my money back?
Contact the distributor or clinic named on your paperwork and quote the recall number on this page. They will tell you whether the remedy is a repair, a replacement or an instruction update.
Where was the recalled product sold?
It was distributed nationwide.
Related recalls
-
Class II · 25 days ago
Abiomed, Inc. Automated Impella Controller with the below product descriptions recall: Retrospective report of issues remedied in 2019 with firmware and software
Retrospective report of issues remedied in 2019 with firmware and software patches: 1. Defects in optical bench (OPB) firmware V2.19, which affected…
-
Class I · last month
Abiomed, Inc. Automated Impella Controller with the below product descriptions recall: The potential for purge cassette recognition issues due to the failure
The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
-
Class I · last month
Abiomed, Inc. Impella Optical Controller with the below product descriptions recall: The potential for purge cassette recognition issues due to the failure
The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
-
Class I · last month
Abiomed, Inc. Impella Optical Controller, Automated Impella Controller recall: The potential for purge cassette recognition issues due to the failure
The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
-
Class I · 3 months ago
Abiomed Impella CP Set, Pump Set Code 0048-0003 recall: A potential for introducer sheath leakage from under the sheath cap and along
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These…
-
Class I · 3 months ago
Abiomed Impella CP Introducer Kit 14 Fr, 13 cm recall: A potential for introducer sheath leakage from under the sheath cap and along
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These…