Roche Diagnostics Operations, Inc. recalls
2 recalls filed under Roche Diagnostics Operations, Inc.. The most recent was published March 11, 2026.
-
Class II · 8 months ago
Roche Diagnostics Operations, Inc. cobas pro integrated solutions with cobas c 503 recall: Software defect, which allows the system to accept erroneous
Software defect, which allows the system to accept erroneous, non-monotonous calibrations for Spline-type assays. Cobas pro integrated solutions is…
-
Class II · 9 months ago
Roche Diagnostics Operations, Inc. The Elecsys Anti-Tshr immunoassay is a three-step recall: Issues identified: 1) Results vary by Instrument Platform
Issues identified: 1) Results vary by Instrument Platform; 2) Calibrator Lot-to-Lot Variability. Issues may lead to delayed or incorrect diagnosis…
About this page
Roche Diagnostics Operations, Inc. appears here because a federal agency named it as the recalling firm or the brand on the label. A recall is not a judgement about everything the company sells — it covers only the lot codes and models listed in each notice.