Beta Bionics, Inc. recalls
3 recalls filed under Beta Bionics, Inc.. The most recent was published June 24, 2026.
-
Class II · 4 months ago
Beta Bionics, Inc. iLet Bionic Pancreas, REF: BB1001 recall: The reason for the recall is: it was found in version 3.3.6 of the iLet ACE
The reason for the recall is: it was found in version 3.3.6 of the iLet ACE Pump firmware and all previous releases that a screen unlock failure…
-
Class II · 6 months ago
Beta Bionics, Inc. iLet Bionic Pancreas, REF: BB1001 recall: insulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous
insulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous Glucose Monitoring (CGM) G7 sensor may result in delayed CGM reading…
-
Class II · 6 months ago
Beta Bionics, Inc. iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision Software recall: it was found that in versions 1.3.7, 1.4.2, and 1.4.3
it was found that in versions 1.3.7, 1.4.2, and 1.4.3, the Lock Screen and Limited Access Passcode Screen on the iLet graphical user interface (GUI)…
About this page
Beta Bionics, Inc. appears here because a federal agency named it as the recalling firm or the brand on the label. A recall is not a judgement about everything the company sells — it covers only the lot codes and models listed in each notice.