Alere San Diego, Inc. recalls
4 recalls filed under Alere San Diego, Inc.. The most recent was published October 7, 2026.
-
Class II · 18 days ago
Alere San Diego, Inc. Fisher Scientific Sure-Vue Urine hCG Strip recall: hCG rapid tests may detect low levels of hCG below the stated assay
hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.
-
Class II · 18 days ago
Alere San Diego, Inc. McKesson Consult hCG Urine Cassette, Model/ Part: FHC-A102/5001 recall: hCG rapid tests may detect low levels of hCG below the stated assay
hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.
-
Class II · 18 days ago
Alere San Diego, Inc. Fisher Scientific Sure-Vue Urine hCG recall: hCG rapid tests may detect low levels of hCG below the stated assay
hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.
-
Class II · 18 days ago
Alere San Diego, Inc. Fisher Scientific Sure-Vue Serum/Urine hCG recall: hCG rapid tests may detect low levels of hCG below the stated assay
hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.
About this page
Alere San Diego, Inc. appears here because a federal agency named it as the recalling firm or the brand on the label. A recall is not a judgement about everything the company sells — it covers only the lot codes and models listed in each notice.